This summary is based on a special report by ECRI Institute.
Can you consider using the Steris System 1E even without a biological indicator (BI)? Yes, but decide carefully, ECRI Institute advises.
A number of questions have been raised about the Steris System 1E (SS1E), which is marketed by the Steris Corporation for reprocessing heat-sensitive medical devices, such as endoscopes. Most recently, concern has arisen about the fact that no BI is available for use with the SS1E.
Although Steris has a BI under review by the Food and Drug Administration (FDA), ECRI Institute does not know when—or even if—it will be cleared. Many hospitals are uncertain about how this affects the purchase and use of the SS1E for critical devices used in the OR. (The lack of a BI is not an issue if the SS1E will be used only to reprocess semi-critical devices outside the surgical setting.)
Resolving these questions is time-critical because the SS1E is seen as a possible replacement for the Steris System 1. The FDA is requiring Steris to stop supporting the system beyond February 2, 2012.
ECRI Institute has discussed this concern with Steris and infection control experts.
Although no BI is available for SS1E, Steris does have FDA clearance for a chemical indicator. The chemical indicator provides confirmation that the dose of sterilant is above the minimum recommended concentration.
Here are some common questions.
Does use of the SS1E without a BI pose a safety risk for patients?
No. We have seen no evidence that there is unreasonable risk in using the SS1E to reprocess devices intended for critical use, provided it is used with device models validated by the manufacturer or Steris for processing in the SS1E and that it is used in compliance with the use instructions.
Is there reason to be concerned about not using a BI with the SS1E?
Yes. For critical devices, routine use of a BI—weekly or daily—for sterilizer efficacy testing is common practice in the US and is recommended by AORN, the Association for the Advancement of Medical Instrumentation, and the Centers for Disease Control and Prevention.
Given these recommendations and the fact that use of BIs during sterilization is common practice, you need to carefully consider the risk management and liability implications of using the SS1E to sterilize critical instruments.
Should you consider the SS1E, even without a BI?
Possibly, if some critical devices (eg, some endoscopes) in your current inventory are not compatible with alternative sterilization methods or if the alternatives have other serious disadvantages.
If possible, ECRI Institute recommends using an alternative sterilizing method other than the SS1E, given the liability and risk management issues described. If all critical devices for which you are considering the SS1E are validated with a method other than SS1E, and workflow and cost do not argue against the alternative, select the alternative.
ECRI Institute also recommends having your Steris System 1 transition team review all of these issues. If you do not have a team, strongly consider including representatives from infection control, reprocessing, clinical engineering, risk management or the legal department, and affected clinical departments in your decision process.
The full report is available to subscribers of ECRI Institute’s Health Devices Alerts www.ecri.org
ECRI Institute. Can you consider using the Steris System 1E even without a biological indicator? Yes, but decide carefully. Health Devices Alerts. February 3, 2011.
Central line-associated bloodstream infections (CLABSIs) declined by 58% in ICUs between 2001 and 2009, the Centers for Disease Control and Prevention reported March 1, 2011.
Some 27,000 lives have been saved and $1.8 million in costs avoided since 2001, the CDC estimates. In 2009 alone, up to 6,000 lives were saved and $414 million in costs avoided.
Cited are prevention efforts by the Agency for Healthcare Research and Quality, the Pittsburgh Regional Healthcare Initiative, and the Michigan Keystone Project.
Expanded prevention efforts are needed in non-ICU settings, especially outpatient hemodialysis centers, the CDC says.