In the never-ending quest to improve patient care, we are always seeking better ways to process medical devices. Many leading experts now recommend transitioning from high-level disinfection (HLD) to sterilization for semicritical items.
Semicritical items are instruments or objects that contact mucous membranes or nonintact skin of a patient, but do not usually penetrate the blood barrier or other normally sterile areas of the body. If possible, these items should be sterilized; if not, they should undergo HLD, which would be expected to destroy all microorganisms except for large numbers of bacterial spores.
The use of HLD for semicritical items is based on the Spaulding Classification System devised more than 50 years ago. Given the types of sterilization and medical devices available back then, it was acceptable to process all semicritical items using HLD. Recent research, however, shows some pathogens are resistant to HLD, medical devices and instructions for use (IFU) are becoming more complex, and new low-temperature sterilization modalities make it possible to sterilize instruments in the same amount of time—or less—as required for HLD.
In 2017, the Association for the Advancement of Medical Instrumentation (AAMI) hosted an Endoscope Stakeholders meeting at which industry leaders and experts discussed infection prevention practices related to the use of HLD for flexible endoscopes. Based on research and discussions at the meeting, it was recommended to transition flexible endoscopes (many of which are classified as semicritical items) from HLD to sterilization. In addition, the AAMI standards and AORN guidelines recommend sterilization over HLD for semicritical items.
A major benefit of transitioning from HLD to sterilization is that sterilized items are dry and packaged, whereas those that have undergone HLD are unpackaged and wet after the process (sidebar, at right). The primary difference, however, is the significantly greater effectiveness of sterilization compared with HLD.
HLD kills all microbial organisms but not necessarily large numbers of bacterial spores, whereas sterilization is a validated process used to render a device free from viable microorganisms (providing a sterility assurance level of 10-6). HLD does provide a 6-log reduction in microbial contamination; however, HLD does not have a sterility claim and therefore does not have a sterility assurance level. Sterilization offers twice that amount at a 12-log reduction.
Sterilization provides quality process monitors that verify the efficacy of the sterilization process. Sterilizers are tested at least daily with a biological indicator process challenge device (BIPCD), which provides a highly resistant challenge to the sterilization process and is a direct measure of the lethality of a sterilization cycle. The BIPCD contains a BI with a large number of highly resistant viable bacterial spores that are most resistant to that method of sterilization.
This method of challenging the sterilization process is so reliable that sterilizer manufacturers use BIPCDs as part of the validation of their sterilization cycles during the half-cycle validation testing.
Sterilization cycles are monitored with physical monitors that record the sterilization parameters. These parameters are reviewed for acceptance before releasing the load, and the records are maintained.
All items that undergo sterilization have both internal and external chemical indicators (CIs). The external CIs show that the package has been exposed to the sterilization process, and distinguish between processed and unprocessed packages. Different types of internal indicators may be used, but they must meet the qualifications described in the ANSI/AAMI/ISO CI standards.
Quality monitors are also available for HLD. These monitors measure either the minimum effective concentration—the minimum concentration of HLD that achieves the claimed microbicidal activity—or the minimum recommended concentration—the minimum concentration at which the manufacturer tested the product and validated its performance.
Currently, there are no standards for these testing devices. When HLD is performed in an automated system, most systems also provide a printout showing the cycle parameters. Manual systems do not have a method of recording the HLD parameters.
Although the main reason to switch from HLD to sterilization for semicritical items is the sterility assurance and greater degree of quality assurance, there are numerous other advantages.
Sterilized items are packaged to be protected from damage and contamination. Because items that have undergone HLD are not packaged, they may easily be contaminated during storage and transport. HLD items often are stored for short amounts of time, typically just days, as per hospital policy. In some cases, an item may need to be processed before use or may not be readily available when needed. In other cases, an item may have to be processed more frequently, which can lead to added wear on the device.
HLD can be performed either manually or by using automated systems such as automatic endoscope reprocessors. Automated HLD processors are superior to manual systems and are recommended for use, according to the AORN Guideline for Processing Flexible Endoscopes. These systems are designed to provide consistency, reliability, productivity, documentation of the cycles, and reduced risk of employee exposure to chemicals. Manual HLD systems present more opportunities for error than automated systems.
Furthermore, technicians using a manual HLD system tend to be inconsistent in how they process medical devices.
Other manual processes include the documentation of a high-level disinfectant as it is opened and poured into a secondary container. Once a bottle of high-level disinfectant is opened, it must be dated to expire (this date is based on the IFU from the disinfectant manufacturer). The secondary container must be clean and closed. All containers used for manual HLD require thorough cleaning between uses.
HLD requires exposure for a specific time and temperature range. In a manual system, ensuring the temperature is reached and maintained is cumbersome, and both parameters require personnel to monitor the process and document this information.
High-level disinfectants are liquid and can result in a chemical spill. Healthcare facilities must ensure that employees are protected from such exposures and provide the necessary personal protective equipment. Whichever HLD system is used, the correct utilities must be available. To reduce the risk of a spill, the HLD system should be located as close as possible to a sink.
Liquid chemical disinfectants are toxic chemicals that can leave a toxic residue and cause patient injury if not properly rinsed. There have been reports of toxic residue injuries, such as toxic anterior segment syndrome (TASS), anaphylaxis, and bowel injury. After HLD, the final rinses should be performed using critical water to prevent contamination. Critical water is extensively treated, usually by a multistep process such as softening, deionization, reverse osmosis, or distillation to ensure microorganisms and inorganic and organic materials are removed.
Transitioning from HLD to sterilization takes planning, communication, and good record keeping. The first step is to involve all stakeholders, ideally by forming an interdisciplinary team with leaders from surgery and sterile processing, a surgeon, and a representative from infection prevention. This team needs to be consulted and communicated with throughout the transition. It is also advisable to inform hospital administration to gain their approval and support.
Next, perform a risk assessment to see if sterilization is feasible and if transitioning to sterilization will improve the quality of medical device processing. Discuss the advantages, disadvantages, and implementation process, considering factors such as:
• level of disease transmission risk of the items processed by HLD vs sterilization
• documented validation testing for a specific HLD or sterilization modality (sidebar, at right)
• storage time and requirements of items processed by HLD vs those sterilized in sterilization packages, ie, amount of labor, whether the item is rapidly available, and whether additional processing is needed that will create more wear on the scopes
• chemical safety for both HLD and low-temperature sterilization (the safety data sheets from these chemicals provide employee and patient safety considerations)
• complexity of the process (the greater the complexity, the higher the likelihood of errors)
• level of assurance that the quality monitors provide for the process that ensures patient safety;
• processing/turnaround time for each process
• labor time and processing time (excluding the automated processes)
• acquisition time variance of obtaining an HLD processed item vs a packaged sterilized item
• damage to items from the process
• residue that may need to be addressed when switching from HLD chemical to another chemical process.
When using a new sterilization modality, review the packaging and quality monitors to ensure they are validated for that type of sterilization. Sterile processing staff must be trained on the sterilization system, new packaging, and quality monitors. Staff who use the medical device should be informed of any changes before they occur.
When transitioning from HLD to a sterilization method, there may be HLD residue on the device. Check with the medical device manufacturer and sterilizer manufacturer, and if there is a risk of residue, find out how to avoid or eliminate the residue.
As the validations are confirmed and recorded for reference, inform the interdisciplinary team of the progress. Moving from an HLD process to sterilization may require a containment device or instrument tray to securely hold the item to prevent damage and to completely expose all surfaces of the item to the sterilant.
If a new type of containment device or instrument tray is used to package the item, it also must be validated for whatever sterilization process is being used, and the cleaning instructions must be reviewed and communicated to sterile processing staff. When transitioning an item to sterilization, the IFU may require special attachments or positioning. Sterilization qualification testing for an item may be performed.
The preparation instructions in sterile processing will need to be changed. For example, some items may require a special cap or adaptor. It may be helpful to have pictures to illustrate these kinds of changes. If an instrument tracking system is used, it will need to be updated and changed to sterilization as it occurs.
All changes involving surgery need to be communicated to the surgical staff. The changes that most impact their functions are:
• Storage. Packaged sterilized items are often stored in different locations than those that undergo HLD; therefore, all staff members need to be aware of the stocking location change.
• Packaging. Users must be aware of the packaging so they can be prepared to aseptically open the package.
• Turnaround time. This is important because it enables the surgical staff to effectively schedule cases.
• Labeling. Users must know which description is being used on the package (this is helpful when searching for the item).
• Quality monitors. Users need to know how to identify an item that has successfully undergone sterilization.
Be sure to keep the infection prevention department head and committee informed as medical devices transition from HLD to sterilization.
As new medical devices are purchased, the product review should include the sterilization modality. Review the IFU for sterilization information. Some semicritical items that do not have an available sterilization modality listed in their IFU may be exempt from the purchasing review. However, there may be some essential semicritical medical devices that lack the sterilization validation claim or a method to sterilize them.
Develop a policy and procedure with the interdisciplinary team that clearly states sterilization is the preferred method of treatment for semicritical items whenever sterilization is available.
This policy will support the requirement of a sterilization claim for further purchases of semicritical items and should be supported and approved by the infection prevention committee and administration.
Medical devices have become more complex over time, and some pathogens have become resistant to HLD. Using HLD to process semicritical items posed fewer risks in the past than it does today, and providing quality patient care now requires moving to sterilization modalities that are more reliable.
Sterilized semicritical items are provided in a ready-to-use state that offers greater assurance that they have undergone processing and are safe for patient use. Leading sterilization experts agree that semicritical items should undergo sterilization whenever possible. There is not a specific target date for this transition to occur, but rather a gradual implementation. ✥
Susan Klacik, BS, CRCST, ACE, CIS, FCS, is clinical educator for IAHCSMM (International Association of Healthcare Central Service Materiel Management) and a representative to AAMI. She also serves on AORN’s Guidelines Advisory Board.
AORN. Guideline for cleaning and processing flexible endoscopes and endoscope accessories. Denver, CO: AORN, 2019. www.aorn.org.
AORN. Guideline for manual chemical high-level disinfection. Denver, CO: AORN, 2019. www.aorn.org.
AORN. Guideline for sterilization. Denver, CO: AORN, 2019. www.aorn.org.
Association for the Advancement of Medical Instrumentation. Chemical sterilization and high-level disinfection in health care facilities. ANSI/AAMI ST58:2013. Arlington, VA: AAMI, 2013. www.aami.org.
Association for the Advancement of Medical Instrumentation. Comprehensive guide to steam sterilization and sterility assurance in health care facilities. ANSI/AAMI ST79:2017. Arlington, VA: AAMI, 2017. www.aami.org.
Association for the Advancement of Medical Instrumentation. Flexible and semi-rigid endoscope processing in health care facilities. ANSI/AAMI ST91:2015. Arlington, VA: AAMI, 2015. www.aami.org.
Association for the Advancement of Medical Instrumentation. Strong Evidence for Sterilization of Endoscopes Presented at Stakeholder Meeting. https://www.aami.org/newsviews/newsdetail.aspx?ItemNumber=5243.
Food and Drug Administration. Design of endoscopic retrograde cholangiopancreatography (ERCP) duodenoscopes may impede effective cleaning: FDA Safety Communication. March 4, 2015. http://wayback.archive-it.org/7993/20170722213105/ https://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm434871.htm
Food and Drug Administration. Gastroenterology-Urology Devices Panel of the Medical Devices Advisory Committee Meeting, May 14-15 2018. https://www.fda.gov/advisory-committees/advisory-committee-calendar/may-17-2017-gastroenterology-and-urology-devices-panel-medical-devices-advisory-committee-meeting.
Ofstead C L, Quick M R, Wetzler H P, et al. Effectiveness of reprocessing for flexible bronchoscopes and endobronchial ultrasound bronchoscopes. Chest. 2018;154(5):1024-1034.
Ofstead C L, Heymann O L, Quick M R, et al. Residual moisture and waterborne pathogens inside flexible endoscopes: Evidence from a multisite study of endoscope drying effectiveness. Am J Infect Control. 2018;46(6):689-696.
Rutala W A, Weber D J. Disinfection and sterilization: An overview. Am J Infect Control. 2013;41(5):S2-S5.
Rutala W A, Weber D J. Gastrointestinal endoscopes: A need to shift from disinfection to sterilization? JAMA. 2014;312(14):1405-1406.