New recommendations for preventing surgical fires seek to change a long-standing practice—open delivery of 100% oxygen to the face during head, face, neck, or upper-chest surgery.

The new advice from the Anesthesia Patient Safety Foundation (APSF) and ECRI Institute includes:
Securing the airway prevents oxygen-enriched gases from accumulating under the surgical drapes. Oxygen-enriched atmospheres contribute to the majority of reported surgical fires, the recommendations note.
There are limited exceptions for open oxygen delivery, such as carotid artery surgery, neurosurgery, and some pacemaker implantations where patients need to be able to speak during the procedure.
“These are significant recommendations that are intended to change clinical practice that has been in effect for 30 or more years,” says Mark Bruley, ECRI Institute’s vice president of accident and forensic investigation, a national expert on surgical fires who is on the committee that developed the recommendations.
For surgery in locations not near an oxygen source, such as the abdomen, groin, legs, and hands, open delivery of oxygen can be used, but the risk of fire is always present, the recommendations note.
The new advice expands on a 2008 practice advisory on OR fires from the American Society of Anesthesiologists.
Preventing surgical fires is a team effort. A good way to assess the fire risk for each case is to do a fire-prevention time-out right before the procedure, Bruley advises.
“A surgical fire prevention time-out shouldn’t take more than a few seconds,” he says. “If there are risks, you can consider the appropriate means of prevention.”
Christiana Care, a health system based in Newark, Delaware, has included a fire risk assessment with the Universal Protocol for surgical site verification for more than 4 years (sidebar, p 21). The assessment scores fire risk from 1 (low risk) to 3 (high risk), with 1 point each for:
A score of 3 means a high-risk fire protocol is initiated.
“The fire assessment just takes seconds—our staff is very used to doing it,” says Judy Townsley RN, CPAN, director of clinical operations/perioperative services.
“Because the fire-risk assessment is embedded in our process, this is really a confirmation,” adds Mary Cay Curran, RN, MSN, CAPA, perioperative manager for clinical standards. OR teams generally know which procedures are high risk and plan ahead. The fire risk assessment is documented as part of the electronic record.
New and updated resources for fire safety education reflect the new recommendations. ?
Christiana Care’s fire safety tools are available at www.christianacare.org/home-page.cfm
American Society of Anesthesiologists. Practice advisory on prevention and management of operating room fires. Anesthesiology. 2008;108:786-801. www2.asahq.org/publications/pc-303-4-practice-advisory-for-the-prevention-and-management-of-operating-room-fires.aspx
Joint Commission. Sentinel event alert. Issue 29, June 24, 2003. www.joint-commission.org
Scoring fire risk for surgical patients. OR Manager. 2006(1):19-20.


Routine protocol
1. Fuel
2. Ignition source
High-risk protocol (includes all of the routine protocol)
Source: Christiana Care, Newark, Delaware. Reprinted with permission.