January 1, 2013

Scope storage: Don’t get hung up on a number

By: OR Manager
PDF
Share

How long can a flexible endoscope be stored before it needs to be reprocessed for use in a patient? Guidelines differ, raising questions about the appropriate storage or “hang time.”

Evidence is limited. What’s the best course? How do accreditation surveyors assess this?

Though infections from GI endoscopes are rare, estimated at about 1 in 1.8 million procedures, contaminated scopes are linked to more health care-associated infection outbreaks than any other medical device, according to the Centers for Disease Control and Prevention.

All of the published incidents of pathogen transmission in GI endoscopy are linked to the failure to follow cleaning and sterilization/disinfection guidelines or defective equipment, notes the 2011 Multisociety Guideline on Reprocessing Flexible GI Endoscopes.

 

Improper cleaning and reprocessing

In discussing hang time, don’t miss the real reason infections are spread—improper scope cleaning and reprocessing.

The most critical aspect of preventing transmissions is to be sure the whole process is followed correctly, stresses Kathryn Snyder, BSN, MM, RN, CGRN, endoscopy/bronchoscopy/motility manager at the University of Virginia (UVA), Charlottesville.

She offers a few reminders:

• Is your staff up to date on the latest society guidelines and manufacturers’ instructions for endoscopy equipment?

• Are all reprocessing steps followed all of the time?

• Does the organization provide the resources to ensure processes are performed correctly?

• Do endoscopy technicians receive annual training and competency validation?

• Is documentation complete and consistent for endoscope reprocessing throughout your organization?



Guidelines on hang time

Two major guidelines differ in their recommendations on storage for flexible scopes based on the same 3 studies (sidebar, p 20):

• AORN advises reprocessing scopes before use if unused for more than 5 days.

• The Multisociety Guideline from the American Society for Gastrointestinal Endoscopy and Society for Healthcare Epidemiology of America says the issue is unresolved and data is insufficient, adding that reuse within 10 to 14 days of high-level disinfection appears safe.

The Society of Gastroenterology Nurses and Associates (SGNA) standards, revised in 2012, refer to the Multisociety Guideline, saying the issue “warrants further data and research.”

 

Making an informed choice

In considering hang time, managers need to review the evidence and make an informed decision appropriate to their organization, advises Cindy Taylor, RN, BSN, MSA, RN, CGRN, nurse manager of GI endoscopy/ bronchoscopy at Hunter Holmes McGuire VA Medical Center, Richmond, Virginia.

“I don’t think there is a right or wrong answer, as long as there is a rationale to back up the decision that is supported by the literature, the standard of care, and society guidelines,” she says.

“Be sure your policy is attainable,” she adds. “Better to not have a policy than to have one and not follow it.”

Some issues to keep in mind:

• GI endoscopes must be properly cleaned and at a minimum subjected to high-level disinfection (HLD).

• Consult with your physicians and infection prevention experts on the proper process for endoscopes used in immunosuppressed patients or in sterile regions such as the biliary tree, pancreas, or peritoneal space.

• If endoscopes are turned over frequently, storage time may not be an issue.

• Keep in mind that in the studies of storage time, the types of organisms cultured from endoscopes after storage were primarily nonpathogenic skin bacteria.

 

The VA’s policy

The Veterans Health Administration currently follows a directive to reprocess unused scopes after 12 days of hang time, Taylor notes.

The hang time is documented:

• using a printout from the reprocessing machine

• keeping the printout in a plastic sleeve attached to the scope by a beaded chain

• scanning reprocessing information into each patient’s medical record, including the HLD parameters, date reprocessed, person who reprocessed the scope, and the reprocessing machine number.

Immediately prior to the scope’s use, the plastic sleeve is removed, and the reprocessing information is verified by a nurse or technician.

“This has become part of our time-out before the procedure,” says Taylor.


 

 

Practice at UVA

UVA is considering adopting a 2-week storage time for flexible scopes, says Snyder. Storage time will be tracked by:

• using a standardized form to document the data and time endoscopes were reprocessed and kept on file for 3 years

• tagging each scope with the date and time it was reprocessed

• removing the tag just prior to the scope’s insertion in the next patient.

“The idea is that you never use a scope without taking the tag off,” she says. “And you take the tag off immediately before insertion, not when you are setting up the scope.” That is in case a physician decides to use a different scope at the last minute.

 

When surveyors visit

A surgeon surveyor from the Joint Commission asked about hang time in a 2010 visit to Taylor’s facility. “He just wanted to know if we had a policy,” she says.

At UVA, surveyors did not ask about hang time during recent inspections by the Joint Commission and Centers for Medicare and Medicaid Services (CMS). But that experience doesn’t necessarily apply to others, Snyder cautions. Surveys vary by state and surveyor. ❖

 

References

AORN. Recommended practices for cleaning and processing flexible endoscopes and endoscope accessories. Perioperative Standards and Recommended Practices. Denver CO: AORN, 2012.

Petersen B T, Chennat J, Cohen J, et al. Multisociety Guideline on Reprocessing Flexible GI Endoscopes: 2011. Infect Cont Hosp Epidemiol. 2011;32:527-537. www.asge.org/uploadedFiles/Publications_and_Products/Practice_Guidelines/Multisociety%20guideline%20on%20reprocessing%20flexible%20gastrointestinal.pdf

Rutala W A, Weber D J. Guideline for Disinfection and Sterilization in Healthcare Facilities. Atlanta GA: CDC, 2008.

Schembre D B. Infectious complications associated with gastrointestinal endoscopy. Gastrointest Endosc Clin N Am. 2000;10:215-232.

Society of Gastroenterology Nurses and Associates. Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes. Chicago IL: SGNA, 2012. www.sgna.org