The Joint Commission announced in June it will look beyond flash sterilization to take a broader view of sterilization. Surveyors will scrutinize the entire process rather than just focusing on which sterilization method is selected.
“If a complete and effective process of sterilization is used, it will be considered an effective sterilization method,” the commission said in a June 15 update.
According to the update, surveyors will review the critical steps of disinfection and sterilization, including:
One theme—manufacturers’ instructions. The Joint Commission says surveyors will ask the staff for manufacturers’ instructions for cleaning and decontamination and will ask how the staff is following the instructions. They will want to see that steam sterilization of all types, including flash, meets the parameters set by the manufacturers of the sterilizers, of wrap or packaging, and of instruments themselves.
Sterilization experts say they’re pleased the Joint Commission plans to look at all steps in sterilization but find some aspects of the update confusing. One issue is the description of flash sterilization as a cycle of “3 minutes at 270°F at 27 to 28 lbs of pressure.”
The description doesn’t refer to dry time, differentiate between gravity and prevacuum cycles, nor mention that flash-sterilized items are intended for immediate use, says Rose Seavey, RN, MBA, CNOR, CRCST, CSPDT, an independent consultant and former director of sterile processing.
Reconciling manufacturers’ instructions is another concern. It’s not always possible to reconcile instructions for sets that require extended cycles, which have become increasingly common, says Cynthia Spry, RN, MA, MSN, CNOR, an independent consultant who chaired the working group that developed the comprehensive steam sterilization standard for the Association for the Advancement of Medical Instrumentation (AAMI).
Sterilizer manufacturers can’t provide instructions for some extended cycles because they fall outside the range of those that can be cleared by the Food and Drug Administration (FDA).
“What’s a user to do when it’s impossible to meet these requirements?” Spry asks.
AAMI recommends performing product testing for these sets, but many facilities aren’t familiar with that, she notes.
Reconciling instructions will be a “horrendous job,” agrees David Narance, RN, BSN, CRCST, manager of sterile processing for MedCentral Health System in Mansfield, Ohio.
He makes sure to keep binders with instructions on hand for the staff— and makes sure they get used. He expects surveyors will ask about some of the more complex equipment, such as power drills and back retractors.
AORN “applauds and supports” the Joint Commission’s effort to begin reviewing the whole sterilization process, Ramona Conner, RN, MSN, CNOR, AORN’s manager of standards and recommended practices, told OR Manager.
She added that AORN is “considering how we can open a conversation and talk further about how to clarify the statement.” This would include terms used and how surveyors plan to implement the update.
She said AORN had discussed sterilization issues with the Joint Commission last year but was not asked to review or comment on the update before it was released.
The broader focus means OR managers “will have to educate everyone in the OR and ensure all of the critical steps are met,” Seavey says.
They will need to make sure sterilizers located in the OR are subject to biological monitoring weekly and preferably daily, as recommended by AAMI.
“The sterilizers in the OR will need to be capable of running multiple types of cycles. Every type of cycle used in each sterilizer will have to be challenged with a BI and with a Bowie-Dick type indicator for dynamic air-removal sterilizers,” she says.
Managers will need to prepare the staff to respond to questions from surveyors. Spry says possible examples are:
Seavey urges managers to be well versed in the AORN and AAMI standards and recommended practices. The Joint Commission refers to the Guideline for Disinfection and Sterilization in Healthcare Facilities from the Centers for Disease Control and Prevention (CDC) and Healthcare Infection Control Practices Advisory Committee (HICPAC).
The Joint Commission update is at www.jointcommission.org/Library/WhatsNew/steam_sterilization.htm
Association for the Advancement of Medical Instrumentation. Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities. ANSI/AAMI ST79:2w006. Arlington, VA: AAMI, 2006, 2008. www.aami.org
Rutala W A, Weber D J, et al. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008. Atlanta: CDC, 2008. (CDC/HICPAC guideline) www.cdc.gov/ncidod/dhqp/pdf/guidelines/Disinfection_Nov_2008.pdf
Association of periOperative Registered Nurses. Perioperative Standards and Recommended Practices. Denver: AORN, 2009.
Spry C. Food for thought: Flash sterilization. OR Manager. 2008;24:18,20. www.aorn.org
The Joint Commission says surveyors, among other activities, will: