According to the Centers for Disease Control and Prevention, about one in 31 hospital patients has at least one healthcare-associated infection (HAI) on any given day. To help mitigate HAIs, implementing infection control procedures and policies, frequently performing hand hygiene by all staff and patients, and keeping the healthcare environment and equipment clean are key. One clinical space that continues to help ensure that infection prevention methods for both patients and staff are met is Ambulatory Surgery Centers (ASCs). ASCs implement a variety of strict infection prevention protocols at their facilities and often establish a central sterile department or location.
The chain of infection is a prime example of why central sterile is important for ASCs. The chain of infection is a series of links that must be connected for an infectious agent to have the ability to transmit from one host to another. These links include infectious agents, reservoirs, portals of exit, mode of transmission, a portal of entry, and a susceptible host. It is the job of a healthcare professional to break at least one, if not all, of the links to prevent transmission.
Infectious agents can be eliminated through cleaning, disinfecting, and sterilizing, as well as by wearing the proper personal protective equipment (PPE) and appropriately changing and discarding PPE. Additionally, air purifiers and disinfectors can be used to remove any modes of transmission in the air. Practitioners can also enhance the safety of immunocompromised patients by placing them in different rooms. Performing these steps correctly helps mitigate the risks of HAIs or surgical site infections for patients.
The following steps in the central sterile process help keep ASC facilities safe from infections.
After a procedure is complete, used instruments should be transported back to the decontamination room as soon as possible to prevent any soil from drying on them. According to a comparison study on the efficacy of cleaning methods for ophthalmic microscopic instruments, “the dried microbial biofilm exhibits bacterial resistance to antibiotics, disinfectants, and sterilization agents.” Dried soil makes the cleaning process more difficult; if instruments cannot be properly cleaned, they cannot be sterilized, which in turn increases the chances of surgical site infections and other harms from dried soil.
When transporting used instruments to the decontamination area, the instruments should be moved in a closed container or a leak-proof cart. According to AORN and Occupational Safety and Health Administration (OSHA) guidelines, when transporting sharps, the container must be puncture-resistant, large enough to contain all contents, and labeled with a fluorescent orange or orange-red biohazard label. Proper transportation prevents damage, potential injuries to personnel, or exposure to bloodborne pathogens.
Unsterile instruments not only pose a safety risk for the ASC, but they also have a financial impact. The cost of throwing away instruments varies depending on the size of the ASC and the number of procedures it performs. In addition, not properly sterilizing, handling, and transporting instruments could result in OSHA penalties. Depending on the willfullness of the violation, the corrective actions, the presence of exposed sharps, and the risk of bloodborne pathogens exposure, ASCs could be fined as much as $70,000. Penalties for repeat violations may incur higher fines.
Once used instruments are properly transported to the decontamination area, they must be cleaned and decontaminated. When cleaning the instruments, follow the manufacturer’s instructions for use (IFU) on how to disassemble and clean (manual or mechanical). The IFUs will also detail if the instrument should be submerged, and which type of cleaning solution works best. This will help promote proper cleaning for the soil to be removed from the instruments. It will also help the instruments last longer since they are cleaned using the manufacturer’s IFU.
If manually cleaning instruments, make sure instruments are submerged under water to help minimize any aerosolization. Also, keep the water and detergent at a lukewarm level where the temperature should range between 27˚C to 44˚C. The lukewarm water level helps keep soil from staying on instruments and makes an easier manual cleaning process. When soaking the instruments, it is advised to monitor and document the temperature of the solution. If using mechanical cleaning for instruments, follow the manufacturers’ IFU on how to load, run, and unload instruments. Failure to comply can cause improper cleaning, thus leading to soil remaining on the instruments. If any soil is on the instruments, they will not be able to be sterilized. Proper testing, monitoring, and documenting of the cycles must be completed to ensure cleaning was done correctly.
Depending on the type of autoclave a facility uses, it is imperative to use the proper packaging for each type of autoclave. The packaging must be compatible with the type of sterilizer being used in a central sterile environment. Using the wrong packaging could cause improper sterilization or mechanical malfunctions that can place the staff in harm’s way. If there is a malfunction with the autoclave, an ASC will know which instruments to pull from rotation because of labeling. According to AORN’s 2021 guidelines, when labeling the packages, information should include identification of the sterilizer, type of sterilizer and cycle used, load control number, a critical parameter for the specific sterilization method, and other elements that can identify items if there is an autoclave malfunction or failure.
Items placed in the packaging should not exceed 25 pounds. It has been exhibited that packages that exceed that weight have difficulty drying, unless the facility uses longer drying times. When wrapping, ensure packages are secure, allowing room for the sterilization process. According to the Association for the Advancement of Medical Instrumentation, there are two acceptable wrapping methods: sequential double-wrapping and simultaneous double-wrapping. The diagonal fold is recommended when wrapping individual small to medium instrument trays, and the square fold is recommended when wrapping large instrument trays.
There are six main types of chemical indicators, called types 1 through 6. Every package must have an indicator both outside and inside. The type 1 indicator should be placed on the outside. This shows that the autoclave reached temperature. Type 5 or type 6 indicators should be on the inside, reading multiple parameters including pressure and time. Type 5 or type 6 are the preferred chemical indicators for the inside of the package, but type 3 or 4 can be used as well. Be sure to check that the chemical indicators are placed correctly.
A recent study on the effectiveness of chemical indicators showed that 13.5% of chemical indicators did not change. Further research was needed to see if there was an autoclave malfunction or operator error that caused that 13.5%; however, if everyone is consistent on placement, monitoring, and documenting all elements in central sterile, it will be easy to discern when something is not working correctly.
The type 2 chemical indicator, also known as a Bowie-Dick test, should be used daily for any gravity displacement steam sterilizers. According to AORN guidelines, running the test before the first load each day and after the warm-up cycle is complete is suggested. The Bowie-Dick test does not replace any biological indicator testing. When using biological indicator testing, use the correct spore test with the type of autoclave and follow the manufacturer’s IFU on how to load, run, and read the test. AORN suggests running a spore test daily, but at minimum once a week on every autoclave in central sterile.
After the cleaning and packaging is complete, the last step in central sterile is sterilization. The facility must always clean and sterilize any reusable semi-critical and critical instruments. Like the other machines and items used throughout central sterile, always follow the IFUs on how to load, run, monitor, unload, and clean the autoclaves. Cycle and drying times must be done correctly. If anything is taken out before the complete cycle, there is no guarantee that the instruments are completely sterilized before being used on another patient. Many pathogens can live on a surface for long periods. For instance, Hepatitis B and Hepatitis C are a few of the pathogens that can remain on a surface for days and transmit easily if not killed correctly. Another indication that sterilization was not completed for steam sterilizers is if packages come out wet after the full cycle is complete. If this happens, a series of investigations must be done to find the error in the sterilization process.
Determine how loaded the autoclave is and if the packages stayed in the full cycle; then, run biological tests to confirm if the autoclave is working or not and see if there are any mechanical errors such as clogged drains. If the error is found, run a series of biological tests to confirm that the autoclave is fixed and rerun packages that were running before the error occurred. If it is not fixed, remove the autoclave and packages from rotation and seek service. To maintain the safety of the patients that the instruments are being used on, as well as the staff using the instruments, it is important to monitor and document every load. If there is any doubt that something is not properly working, pull all instruments from the failed cycle out of rotation and begin tests to diagnose the problem.
Lastly, other solutions can help maintain the sterility of instruments in an ASC. One solution is the reprocessing of single-use devices. Facilities cannot reprocess single-use devices. However, using an FDA-cleared third-party reprocessing manufacturer, a facility can guarantee the proper sterilization and inspection of single-use devices. Reprocessing can help limit costs in the central sterile department, as well as return the devices to a state as if they were new from the manufacturer. Finding ways to mainstream central sterile while keeping proper infection control can help keep processes and procedures consistent.
ASCs must always remember to follow the IFU, monitor and document materials, run tests to make sure all equipment is working, and follow the steps laid out by guidelines from the appropriate regulatory bodies. By practicing all these steps and prioritizing having a sterile environment, the safety of both patients and staff can be achieved.