Asking who, what, why, when, where, and how—otherwise known as the “5 Ws and an H”— is a time-tested way for writers and researchers to ensure comprehensive coverage of any topic. Here, we apply this framework from the perspective of sterile processing department (SPD) professionals seeking to start a water quality management program.
We can all imagine a sterile processing (SP) technician opening the steam chamber door only to find wet packages or even white, chalky residue on the packages and chamber walls. These are clear indicators of poor steam quality or of unfiltered water entering the manual sink or sterilizer chamber, potentially compromising the safety of surgical patients.
What should be done about it? Where should samples come from, and how should they be taken? Answers to these and the other “Ws” can be found in the ANSI/AAMI ST108:2023 water quality standard for medical devices. However, the science of water management historically has been left to facility engineers. The following aims to demystify this potentially daunting standard by addressing SP leaders’ most common concerns, beginning with the most fundamental question:
Poor water quality can compromise the most stringent sterilization processes. Defined by the Centers for Disease Control and Prevention (CDC) as containing elevated microbial levels or incorrect chemical composition, poor-quality water can lead to residue buildup, corrosion, and, ultimately, compromised instrument functionality. Contaminated surgical instruments also can result in potentially serious hospital acquired infections (HAIs). A prime example is the nearly threefold increase in Legionnaire’s disease cases in US hospitals from 2000 to 2009.
As documented by Decker and Palmer, sections of plumbing without active flow (“dead legs”) create an ideal environment for water to stagnate, leading to the buildup of scale, sediment, and biofilm. This is an ideal environment for supporting colonization by resilient, gram-negative pathogens (characterized by a protective outer shell) known to cause HAIs. In addition to Legionella, other examples include Pseudomonas species and nontuberculous mycobacteria, a chlorination-resistant organism often responsible for pseudo-outbreaks after tap-water rinsing. Opportunistic molds like Aspergillus and Fusarium also have been found in hospital water and plumbing fixtures.
Decker and Palmer point to hand hygiene as the most critical measure for preventing Gram-negative HAIs, but adhering to best practices makes little difference with a contaminated faucet, aerator, or drain. Overall, the authors call for a multifaceted approach combining engineering controls, hygiene practices, and clinical strategies. It follows that maintaining water quality is the responsibility of everyone involved in patient care.
However, a thorough understanding of the science of water—and the standardized framework provided by ST108 terms and guidelines—is particularly critical for those tasked with sterilizing surgical instruments. Armed with this knowledge, they can adopt a proactive approach to monitoring and controlling parameters that not only keeps patients safe, but also sets an example for a facility-wide culture of excellence in healthcare delivery.
Every SP professional should seek to better understand water quality. However, developing one or two “super users” can be a springboard toward greater department-wide expertise. As the department’s designated expert, the super-user should take responsibility for educating other staff members and ensuring compliance with the ST108 standard. This person also should lead the team in gathering water samples, documenting results, determining the cause of wet loads, and addressing specific problems, such as black-stained instruments returned from the OR team. Although ST108 says nothing about super users, AORN also endorses this approach as a strategy for meeting or exceeding water quality standards and preparing departments for future accreditation surveys.
Regardless, developing a comprehensive and effective water quality management system is likely to require the expertise of non-SP personnel, such as facility engineers, steam experts, or third-party consultants. For example, a third party may be better suited for addressing issues with dead legs, old piping, or water main breaks. Integrating ST108 into processes specifically designed for the SPD also requires clear communication with biomedical engineering departments, which have historically managed water quality. Communication also should extend to the entire interdisciplinary team involved in developing the water system. The AORN Journal recommends setting clear expectations for other stakeholders to provide regular, detailed reporting of test results, including actionable insights and recommendations, as well as immediate notification of any deviations from standard parameters.
The ST108 standard defines three types of water, each with a specific purpose:
• Utility water is generally supplied by the facility’s municipal source. It is suitable only for general purposes and initial stages of the process, such as pre-cleaning to remove gross soil and debris. Potential impurities make it unsuitable for final rinsing or steam generation.
• Critical water meets the stringent purity standards required for a final rinse before instruments enter the sterilizer. The standard outlines specific, acceptable minimums for various minerals, endotoxins, and microorganisms. For example, hard water is prone to leaving deposits, so calcium and magnesium levels must be less than 1 ppm.
• Steam water—water destined for steam-generating autoclaves—also must be purified to eliminate the risk of residues or stains that could compromise performance and safety.
ST108 also outlines specifics on what to inspect and what issues to document. For example, visual inspections of mechanical equipment and sterilizers can reveal residue, staining, scaling, and discoloration. Particular focus should be devoted to critical components such as the spray arms in mechanical washers and the internal chambers of various equipment, including washers, automated endoscope reprocessors, ultrasonic cleaners, and sterilizers.
Visual inspection of equipment should occur daily. The ST:108 standard also specifies a strict schedule for quality checks of incoming feed water.
For example:
• PH measuring should be conducted monthly for critical water, and quarterly for utility and steam water.
• Conductivity, total alkalinity, and total hardness should be measured monthly for critical water, and quarterly for utility and steam water.
• Point-of-use bacterial testing should be conducted monthly for critical water, and quarterly for utility water (this testing is not applicable for steam, which, in this context, provides high enough pressures and temperatures to eliminate bacteria).
For municipal water, the local water authority’s annual report provides necessary information on complying with Environmental Protection Agency standards, identifying temporary excursions, and ensuring sufficient filtration to eliminate particulates.
Feed water sampling should be conducted at key stages of the process. These include the beginning and the end of the water distribution loop, after softening, after reverse osmosis, and prior to final rinsing.
For utility water, check points of delivery to washers, decontamination sinks, or nearby plumbing. Sampling from the exit of storage tanks is also crucial to verifying water quality post-storage. Steam condensate should be collected at the delivery point for the sterilizer or steam filter. For critical water, getting as close as possible to the point of contact with a medical device can help ensure a representative sample.
Collecting samples prior to water treatment system disinfection is imperative to establish a reliable baseline for quality. The ST:108 standard also specifies the entire system—including storage tanks—should be drained and flushed to remove residual disinfectants.
From there, ensure outlet taps run for at least 60 seconds before collection. Flushing out stagnant water, as well as biofilm that tends to accumulate in less frequently used areas of the distribution system, ensures a representative sample that accurately reflects overall water quality.
The standard also specifies the use of sterile, endotoxin-free containers to prevent contamination from affecting test results. Avoid containers made of materials such as polypropylene, which can inhibit the detection of bacterial endotoxins.
Collected water should immediately be stored in ice or at 4°C to prevent bacterial replication and preserve sample integrity until they can be delivered to the laboratory for testing.
As we all know in SP, if a task was not documented, it was not performed. In addition to basic information like the date and the name of the inspector, documentation should include test results, observations, and any cleaning, descaling, maintenance or other corrective actions taken in response to deviations from established quality parameters. Follow-up inspections should also be conducted to ensure the issues have been resolved and the equipment is operating correctly.
Although adding more tasks might seem overwhelming given the existing responsibilities of SP professionals, this systematic record-keeping is essential. Among other reasons, documentation is important for ensuring proper reprocessing, and for ensuring that issues are identified and addressed properly. In addition to providing a record of process integrity, proper documentation ensures effective communication between SPDs and facility leadership. Overall, it is essential to support a culture of continuous improvement, and to ensure that patient safety and care standards remain top priorities.
Overall, understanding the complexities of water quality is not just a theoretical exercise. Rather, it is a critical aspect of maintaining patient safety and ensuring the efficacy of medical instrument sterilization. SP professionals play a pivotal role in this endeavor. By embracing the “5 Ws and an H” approach, these teams can effectively navigate the nuances of water quality assessment and management as outlined in the ANSI/AAMI ST108:2023 standard. ORM
—Angela Lewellyn, LPN, CRCST, CER, CHL, is director of development & research at Advantage Support Services. She has a combined 20 years of experience in surgical services as a circulator, surgical technologist, and sterile processing expert.