January 1, 2025

Why active implants demand proactive management

By: Mary A. Marvin, APRN-BC and Jill Teubel, MSN, RN
Share

What happens when a surgeon uses the monopolar instrument set on 30-W coagulation mode to create an upper midline incision in a patient with a pacemaker? Pacemaker function is interrupted, causing a heart block that results in hemodynamic instability—or at least, this is what could happen without taking the necessary precautions. Depending on the situation, this might mean opting for bipoloar instrument instead, switching to cut mode, decreasing generator power, opting for dessication over fulguration, or reorienting the instrument set’s active electrode relative to the pacemaker.

These strategies are not new. Outlined nearly 10 years ago in Current Problems in Surgery, all are means of addressing a problem identified as early as 1986, when Levine et al warned electrocautery could induce “damage to the pulse generator, reprogramming of the pacemaker, changes in the capture threshold, and ventricular fibrillation.” Yet, evidence suggests education among surgeons on this is lacking.

Meanwhile, the stakes are getting higher. As technology advances, surgical patients are more likely to present with not only pacemakers, but also implantable cardio defibrillators, deep brain stimulators (DBS), cochlear implants, or other active devices (ie, implants with integrated sensors) that could be compromised by electromagnetic interference (EMI) from surgical instruments. Some patients may even present with more than one of these devices.

On the instrument side, advancing diagnostic and treatment technologies—MRI, laser, lithotripsy, radiotherapy, proton therapy, hyperbaric chamber, etc—are expanding potential sources of EMI beyond electrocautery tools. Failing to mitigate the risk can not only harm patients, but also incur significant costs due to delayed or cancelled surgeries and potentially even lawsuits.

 

Higher stakes

Data from the authors’ own organization—OSF HealthCare Saint Francis Medical Center, a Level 1 trauma center in Peoria, Illinois—reveal the expanding variety of active implants. In 2021, a manual case count showed 2.4% of surgical patients (1,000 of 40,000 total) had an implant that required presurgical coordination of care. By 2023, the same metric was 4.5%—a 37% annual increase. The team also identified 37 different categories of active implants that may require perioperative intervention—a number likely to grow as new devices come to market.

Electrosurgery—a common means of maintaining hemostasis—can be risky even for patients without active implants. As documented by Martinsen et al, “tissue burns can develop in certain areas around electrical conducting implants (active or passive). The highest risk typically occurs in small, exposed areas in electrical contact with the tissue.”

Whatever the nature of the issue, liability can be costly. A recent, 33-year review of medical litigation identified 139 cases, 114 (82%) of which were operative burns and 25 (18%) of which were surgical fires. The face was the most common site for operative burns, which most commonly originated from a high-energy device. Two patients died. Plaintiff rewards ranged from $82,000 to $518,000, with a median award of $215,000.

Other reasons to address this concern include costs related to unsatisfied patients, frustrated staff, procedures lost to competition, and, most obviously, surgical delays and cancellations. When margins are tight—a median of 2.7% for hospitals, according to a 2018 analysis from Becker’s Hospital Review—even incremental improvements in OR efficiency can have an outsized impact. And according to the International Journal of Surgery, “Surgery cancellation is the leading cause of [OR] inefficiencies.

 

Greater challenges

As documented by the Society of American Gastrointestinal and Endoscopic Surgeons (SAGES), “Surgeons are not required to train on energy-based devices or document their knowledge of safety issues related to their use.” Additionally, researchers wrote, “their understanding of how to safely use the devices has never formally been tested.” In an assessment of whether gastrointestinal surgeons could learn key facts in a half-day course—specifically, Fundamental Use of Surgical Energy (FUSE)—the authors found “31% of surgeons failed to correctly answer the item on how to handle a fire in the [OR]; 31% could not identify the device that does not interfere with cardiac pacemakers; 13% did not recognize that thermal injury can extend beyond the tissue grasped in a bipolar device; and 10% thought a dispersive electrode should be cut to fit a child.”

SAGES published this research more than a decade ago. However, a 2023 study in Surgical Endoscopy revealed the knowledge gap persists at a time when the baby boomers are aging and more patients are presenting with multiple active implanted devices. Other factors further complicate the problem. For example, frequent staff turnovers can cause a trickle-down educational effect that results in missed opportunities for training awareness efforts.

However educated the staff, relying too much on prior knowledge and experience can lead to errors. Implant manufacturers continue to innovate, and new devices could require special considerations and coordination prior to the day of the procedure. Accounting for such considerations requires cross-referencing the instruments planned for use during the case with the specific device type and model that is in situ. For example, if monopolar electrosurgery is planned for use and the patient has a spinal stimulator from manufacturer X, that combination must be looked up for precautions or contraindications.

Overconfidence among surgical staff is particularly risky with one of the most familiar active devices: the pacemaker, implanted in more than 1 million patients worldwide and 200,000 in the US alone every year. Citing the European Heart Rhythm Association (EHRA) consensus, Collins et al note patients with cardiac implantable electronic devices (CIEDs) “demand distinct consideration throughout the perioperative period with a ‘no-one-size-fits-all’ approach for the procedural considerations fundamental to each device type. Individual devices and manufacturers have unique properties and programming.” For example, leadless pacemakers from Abbott—one of two manufacturers of leadless devices—respond to a magnet, whereas those from the other major manufacturer, Medtronic, do not.

Instructions for use (IFU) do not always have answers. For example, all pacemaker manufacturers provide guidelines for best practice magnet placement, but this information is not always found in the IFU. Based on the category of device, grounding pad placement can alter between device manufacturers, with some companies specifically stating not to use whole body grounding pads within their device’s IFU. Overall, extensive preoperative preparation is essential to preventing patient harm, liability issues, and costly day-of-surgery delays and cancellations.

 

Essential preoperative precautions

In addition to addressing education gaps, ensuring surgical nurses and physicians have access to the information they need when they need it requires effective systems for identifying devices prior to the day of surgery. It also requires a means to keep staff updated amid constant change and the introduction of new implant technologies.

At OSF Saint Francis, efforts to achieve these goals began with the preadmission testing (PAT) center. In addition to early identification of patients with existing implantable medical devices, the PAT nurse coordinates care by, for example, obtaining cardiac magnet/rep orders or reminding patients to bring their remotes so their devices can be turned off prior to surgery.

The preadmission testing nurse also documents the device type with pertinent information in the electronic health record, which appears on the surgery schedule so the surgeon and anesthesia provider can plan accordingly. Ensuring care coordinates with manufacturers’ IFUs is essential, so a system is in place to ensure all necessary documentation is prepared in advance and at the fingertips of the entire perioperative team, from the PAT to the postanesthesia care unit.

Staff training is also critical. The plan developed at OSF Saint Francis touches several areas of the hospital, training providers on the importance of mitigating these risks and procedures for handling patients. Reinforcement of the standard procedures over a long period of time helped enact the cultural shift required to build collective awareness of risks and mitigation steps.

Internal review shows that these efforts are helping prevent approximately 500 delays and cancellations and saving an estimated $1.9 million per year, not including avoiding the cost of medical liability cases. However, such savings need not be limited to OSF Saint Francis. By prioritizing the development of a well-structured plan for risk mitigation, perioperative teams can save time and money while ensuring a safe, seamless surgical experience for their patients. ORM

 

—Mary A. Marvin, APRN-BC, became chief quality and safety officer at OpenSurg, Inc after retiring from OSF HealthCare Saint Francis Medical Center in Peoria, Illinois, where she co-created DeviceWise, an app designed to streamline the management of medical devices.

 

Jill Teubel, MSN, RN, co-creator of DeviceWise, is chief innovation officer at OpenSurg and clinical nurse educator at OSF Healthcare.

 

References

Abate S M, Chekole Y A, Minaye S Y, et al. Global prevalence and reasons for case cancellation on the intended day of surgery: A systematic review and meta-analysis. International Journal of Surgery. August 20, 2020.

Bhatia N, El-Chami M. Leadless pacemakers: a contemporary review. Journal of Geriatric Cardiology. April 2018.

Choudhry A J, Haddad N N, Khasawneh M A, et al. Surgical Fires and Operative Burns: Lessons Learned From a 33-Year Review of Medical Litigation. American Journal of Surgery. March 2017.

Collins E, Marcucci M, Liu J C. Surgical Implications of Implanted Cardiac Rhythm Devices. The Journal of Innovations in Cardiac Rhythm Management. January 2023.

Ellison A, Kim C J. 224 hospital benchmarks: 2018. Becker’s Hospital Review. June 25, 2018.

Feldman L S, Fuchshuber P, Jones D B, et al. Surgeons don’t know what they don’t know about the safe use of energy in surgery. Surgical Endoscopy. October 2012

Gillies M A, Wijeysundera D N, Harrison E M. Counting the cost of cancelled surgery: a system wide approach is needed. British Journal of Anesthesia. September 6, 2018.

Nechay T V, Loban K M, Chechin E R, et al. Surgeons know that they don’t know about the safe use of surgical energy: an international study reveals that the knowledge gap persists. Surgical Endoscopy. June 2023.