May 12, 2020

Device safety depends on routine cleaning, maintenance

By: Susan Klacik, BS, CRCST, ACE, CIS, FCS
Share

Routine cleaning and maintenance of the equipment used to process medical devices is important to prevent contamination of medical devices. This article will discuss the required cleaning and maintenance in a sterile processing department (SPD) as well as the guidelines and regulations that inform these practices.

 

Environmental issues

Recommendations for cleaning and disinfecting surfaces in an SPD are based on manufacturers’ instructions for use (IFU), government regulations, and standards and guidelines. SPDs should have policies and procedures that specify standards for cleaning, disinfection, and routine maintenance. An interdisciplinary team with staff from the SPD, environmental services, perioperative nursing, and infection prevention should work together to determine cleaning procedures and frequency based on the IFU, best practices, and regulations.

This interdisciplinary team should identify high-touch areas to be cleaned and disinfected. Evidence has shown that high-touch areas such as control panels, switches, knobs, and handles can harbor contamination and should be cleaned and disinfected on a routine basis to prevent the spread of contamination.

To ensure the department is being properly cleaned, establish a method for evaluating the effectiveness of the cleaning process. Either a qualitative method, such as visual inspection and/or fluorescent marking, or quantitative methods, using cultures or adenosine triphosphate (ATP) monitoring, may be used. The testing time intervals should be based on recommendations from the interdisciplinary team. Results from quantitative methods should be recorded and can be used to provide measurable, objective data to help drive process improvement, encourage compliance, educate personnel, and verify personnel competency.

To avoid the risk of any missed surface or equipment cleaning, responsibility for cleaning should be assigned. In most facilities, environmental services staff are responsible for cleaning floors, walls, ceilings, and vents. The rest of the SPD environment and processing equipment is the responsibility of SP technicians.

Have a written plan to specify who is responsible for each surface and piece of equipment so that nothing is overlooked. The equipment and work surfaces used in medical device processing should be cleaned on a routine basis as prescribed in the policies and procedures.

Develop checklists specifying the equipment or surface to be cleaned, and the cleaning and maintenance schedules. Checklists are a means to standardize procedures, prevent missing surfaces or equipment, and demonstrate compliance with policies and best practices.

 

Routine cleaning

These are areas that should be considered for routine cleaning and/or disinfection in an SPD:

• Ventilation ducts, including air vents and grills and the filter, should be cleaned and changed on a routine basis. The manufacturer’s IFU should provide this information.

• Walls, floors, and ceilings should be routinely inspected for debris and staining. Any defective areas should be replaced or repaired.

• High-touch areas that cannot withstand disinfection should be protected with a barrier covering to prevent the surfaces from becoming a reservoir for microorganisms. For example, a protective covering for a computer keyboard should be removed or disinfected per the IFU.

• High-touch areas, such as doorknobs, control panels, cart handles, and switches, should be cleaned routinely.

• All floors should be cleaned and disinfected with a damp mop each day they are in use.

• Sinks should be cleaned and disinfected at least daily.

• Emergency eye/shower stations should be cleaned and disinfected on a routine basis. This equipment requires weekly testing to ensure its safe operation according to the Occupational Safety and Health Administration’s Eye and Face Protection Standard (29 CFR 1910.133). Both the shower (if available) and eye wash station should be activated for a period long enough to verify operation and ensure that the flushing solution is available. Water temperature should be monitored to ensure it stays between 60° and 100°F. There should be a record of this weekly test.

• Countertops should be cleaned and disinfected at least daily.

• Workstations should be cleaned and disinfected at the end of each shift and whenever they are soiled.

• The service area behind the sterilizers should be routinely cleaned (with areas in need of cleaning identified). The cleaning schedule should be based on the recommendations from the interdisciplinary team.

• Staff changing areas should be cleaned daily, and staff should clean their lockers regularly.

 

Routine maintenance

It is imperative that staff follow the IFU for SPD equipment. The IFU should specify the types of cleaning and disinfectant chemicals to use, routine maintenance requirements, replacement parts (including tubing and filters), inspection, and other maintenance necessities. These also should be included in a checklist, and verification testing should be recorded.

There are some commonalities of all SP processing equipment that are included in standards, guidelines, and regulations. The IFU provides information about which types of cleaning solutions to use, but not necessarily the frequency of use. Listed below are some of the frequencies of use:

• Ultrasonic cleaners. Clean this equipment at least daily whenever the equipment is in use, in accordance with the equipment manufacturer’s IFU; the basin may require daily cleaning. In addition, the equipment should undergo daily performance verification and cavitation testing. If a manifold system with tubing is used for irrigating cannulated instrumentation, test the tubing for fluid flow. The strainer should be cleaned at a minimum each day it is used.

• Washer/decontaminator. Check the spray arms at least daily to ensure they are completely free-turning and that the spray nozzles are not clogged. Clean the strainer at least daily and when there is visible debris. Clean the external surface at least daily and whenever soiled. Perform cleaning verification testing each day and document it each day the equipment is used. Check to be sure the settings are correct, and examine the printout to ensure proper function. Inspect carts and trolleys, and make sure they are inspected and cleaned on a routine basis or in accordance with recommendations from the interdisciplinary team.

• Cart washer. Perform and verify cleaning verification testing each day the cart washer is used. Check the settings to ensure they are correct, and examine the printout to ensure the cart washer’s proper function. Clean the strainer at least daily and when debris is visible. Check the spray arms at least daily to ensure the arms are completely free-turning.

• Leak tester used to test flexible endoscopes. Pressure verification should be performed for each type of leak tester on a daily basis; results should be recorded.

• Borescopes. Clean and disinfect these after each use.

• Automatic flushers. Check with the manufacturer for specific information on filter changes, tubing cleaning and disinfection, and calibration.

• Detergent doser. Clean and disinfect according to the IFU, and routinely verify for correct dose. Verify calibration per the IFU.

• Pass-through window. Clean and disinfect this daily, and check periodically to ensure that this physical barrier completely closes and seals.

• Instrument preparation and packaging workstations. Clean and disinfect these after each shift.

• Heat sealers. Clean these at least daily to remove packaging debris, and check temperature calibration to make sure it complies with the IFU.

• Biological incubators.Check with the manufacturer for cleaning and other maintenance requirements.

• Storage areas. Sterile storage equipment such as shelves or cabinets should be routinely cleaned and documented.

• Automatic endoscope reprocessor (AER). The AER should undergo automated cleaning efficacy verification testing, and the equipment manufacturer’s written IFU should be followed.

• Sterilizers. Inspect and clean sterilizers in accordance with the IFU. Daily sterilizer evaluation requires assessment of the recording charts, printers, printer ribbons, marking pens, and ink; inspection of the door gaskets for nicks and stains; inspection and cleaning of the chamber drain screen to remove lint tape and other small objects from entering the exhaust line; and inspection and cleaning of external surfaces. The chamber should be cleaned weekly.

• Low-temperature sterilizer. Some low-temperature sterilizers require routine filter changes; as always, it is important to review the IFU and include maintenance functions on a checklist. Chambers should be cleaned per the sterilizer manufacturer’s IFU.

• Carts. Sterilizer carts and trolleys should be cleaned routinely as per the schedule determined by the interdisciplinary team. If sterilizer carts and trolleys are not routinely cleaned, dust and lint may accumulate.

• Scope drying/storage cabinets. The interdisciplinary team should develop a routine cleaning schedule, and a cleaning log should be maintained for cleaning the endoscope storage cabinet. Cabinets with filtered air require filter changes on a routine basis; consult the IFU for specifications. If the cabinet has tubing and adaptors, the IFU should be consulted for specific maintenance.

 

Ensuring safety

It is critical to maintain the equipment used in medical device processing to prevent the equipment from becoming contaminated or not functioning properly. By following the manufacturers’ IFU and recommendations from a multidisciplinary team within the facility, OR and SPD leaders can ensure the integrity of their processes and the safety of their instrumentation. ✥

 

References

AORN. Guideline for environmental cleaning. Denver, CO: AORN, 2020. www.aorn.org

 

Association for the Advancement of Medical Instrumentation. Comprehensive guide to steam sterilization and sterility assurance in health care facilities. ANSI/AAMI ST79:2017. www.aami.org.