In October 2019, AORN released its updated Guideline for Sterilization Packaging, which is based on evidence from peer-reviewed literature. This article includes recommendations from that guideline and addresses common misconceptions about sterilization packaging.
Other resources for proper use of sterilization packaging include the Association for the Advancement of Medical Instrumentation (AAMI) standards, the Centers for Disease Control and Prevention’s Guideline for Disinfection and Sterilization in Healthcare Facilities, and the US Food and Drug Administration (FDA), which regulates sterilization packaging manufacturers and clears sterilization packaging for use in healthcare facilities.
The main function of sterilization packaging is to serve as a sterile barrier and maintain the sterilization integrity of the contents. Many factors must be considered when choosing the appropriate type of packaging. Some basic rules of thumb are that the packaging should:
• allow air removal to permit the sterilant to penetrate the package
• resist puncturing and tears
• allow for a complete seal that is tamper-evident
• be free of toxic components and nonfast dyes
• be non-linting
• allow for aseptic presentation.
In addition, packaged items must be clean and free from contaminants, oils, and soils, so the sterilant can have complete surface contact. To prevent these soils from being transferred from the hands of personnel preparing and packaging medical devices, AORN recommends that personnel perform hand hygiene within the hour or wear clean gloves to perform these tasks.
Sterilization packaging must be cleared by the FDA and validated for the specific type of sterilization it is undergoing. Current instructions for use (IFU) for the packaging system should be readily available to staff.
For steam sterilization, not all packaging systems are validated for both dynamic air removal and gravity cycles. Not all packaging systems are validated for immediate use steam sterilization. The IFU can provide this information, and the packaging IFU provides other critical information, such as shelf life testing. Many packaging systems have been validated only for a specific shelf lifetime.
Here are the features to look for in packaging systems.
Paper-plastic pouches should be used for small, lightweight, low-profile items. Heavy metal instrumentation such as drills and weighted speculums could result in inadequate drying and tears in the packaging.
Double packaging with paper-plastic pouches is allowable only if the pouch manufacturer has validated them for that purpose; otherwise, the capability of the package could be exceeded.
When double pouching, two sequentially sized pouches should be used. Because the sterilant is impervious to the plastic side of the pouch, the pouches should not be folded, and they should be placed so plastic faces plastic to permit the entrance and removal of the sterilant.
Placing paper-plastic pouches inside packaged sets is not recommended because doing so may compromise adequate air removal, sterilant contact, and drying. The only exception to this recommendation is if the pouch packaging manufacturer has validated it for this purpose and it is verified by product testing performed by the healthcare facility.
The IFU may contain information regarding lumen restrictions, types of materials to include, weight limits, specific filters to use, stacking allowances of the same sterilization container system, and specific types of sterilization modalities and cycles to use. To avoid damage to the container, only the types of cleaning agents listed in the IFU should be used, and the sterilization container systems should be cleaned before each use. Protective mats, dividers, and sorting pins that are not attached to the tray should be disassembled for cleaning.
The issue of placing loaned instrument sets into rigid sterilization containers is addressed in ANSI/AAMI ST77:2013, Containment Devices for Reusable Medical Device Sterilization. This guidance was developed for medical device manufacturers to help them address issues related to processing, handling, and validating sterilization for medical devices placed in rigid sterilization containers. Medical device manufacturers must ensure that their devices are compatible with the recommended packaging methods and provide written instructions for reprocessing.
According to this standard, loaned sets may be placed into a sterilization container system only if both the sterilization container and loaned set manufacturers have performed the required validation studies. This restriction is necessary in part because of the worst-case volume-to-vent (V-to-V) ratio. This ratio is defined as the interior volume of the sterilization container divided by the total cross-section area of the perforated vent holes. V-to-V ratio is influenced by the sterilization methodology. For example, a dynamic air removal system will have fewer V-to-V ratio concerns compared to a gravity displacement system. When the V-to-V ratio is critical to the container system under consideration, a higher V-to-V ratio presents a more difficult challenge to achieve the desired lethality performance and should be assessed.
Sterilization package labels serve multiple functions, including listing the package contents and serving as a method to track packages’ sterilization processes (and for recall situations). The contents of the package should be clearly and neatly labeled. Packages that are not properly labeled can result in lost instrumentation or damage to the packaging integrity from staff searching through the storage area and handling packages.
Required tracking information on the package label includes content identification, date of sterilization, sterilizer unique identifier, sterilization cycle and load number, and identification of the person responsible for preparing and packaging the item. This information may be handwritten or placed on a load sticker that is then affixed to the package.
Many facilities use instrument tracking systems. If the instrument tracking system uses a label that provides traceability of the sterilizer’s unique identifier, the load stickers or handwritten information are not required.
AORN’s updated Guideline for Sterilization Packaging was based on an assessment of peer-reviewed literature appraised for the quality of the evidence. Highlights of changes include:
Packaging begins with the pre-purchase evaluation of a packaging system. The guideline provides a listing of considerations such as:
• product quality assurance testing results
• compatibility with the intended sterilization method(s) and cycles used within the facility
• requirements for cleaning according to the IFU (eg, laundry for textiles, equipment for cleaning rigid containers)
• requirements for tracking use
• method for tracking use.
The position on the environmental consideration of packaging systems was updated. Being a patient advocate is not limited to just the surgical procedure (human health depends on the environment, and it is important that we protect our environment).
The updated guideline recommends selecting an environmentally preferable sterilization packaging system, if the systems are equivalent in performance.
Evaluation/selection considerations may include the packaging’s reusability and disposal requirements. It is recommended that packaging for sterilization be performed in an area intended, designed, and equipped for sterilization packaging activities, such as the packaging area of the sterile processing department. Users are advised to verify that instruments and other medical devices have been cleaned, inspected, and assembled according to the manufacturer’s IFU prior to packaging.
Colored or tinted tip protectors are recommended for sharp items to protect instrumentation from damage and to protect personnel from sharps-related injuries. Colored or tinted tip protectors make it easier for surgical personnel to see the sharps, which is important on a sterile field (photo, at right). Clear tip protectors are hard to see and can increase the risk for retained surgical items.
The practice of packaging loaned instrumentation is addressed. It is recommended that sterilization packaging information for loaned instrument sets be obtained from the vendor that provides the instrumentation.
Standardized process is recommended. Research shows that standardizing sterilization packaging processes can increase efficiency and help prevent errors (eg, missing or wrong items in a set). Organizing accessories can help users implement standardized procedures.
The use of count sheets inside of instrument sets has been a concern. It is recommended that each healthcare organization determine whether count sheets may be placed in trays. During this review, it is important to consider the limited research regarding the safety of subjecting toners, inks, and various papers to any sterilization method. Chemicals used in the manufacturing of paper, toners, and inks pose a theoretical risk of reaction in some sensitized individuals; however, one research study (referenced in the updated AORN guidelines) concluded that the label and toner ink transferred during sterilization was not cytotoxic and that further study is needed to incorporate a larger sample, various sterilization methods, and instruments of a variety of compositions.
Corner guards for flat wrapped sets are being used in many surgical instrument sets to prevent holes in a wrapper. Such holes may allow microbes to enter a sterilized package and contaminate it. Personnel should check the IFU of the corner protector manufacturer for information regarding the type of sterilization method for which the packaging is validated.
The updated guideline offers two new recommendations for rigid sterilizer containers. It is now recommended that users establish and implement a schedule for routine rigid sterilization container inspection, maintenance, and repair. Also, new information about what to look for in the container’s IFU has been added, such as:
• instructions for inspection and routine maintenance
• recommended types and placement of instruments and other medical devices
• the most challenging location within the package for placement of internal chemical indicators and internal biological indicators.
Selecting the correct sterilization system depends on the medical device and sterilization modality. For this reason, it is important to base packaging selection on the medical device and packaging manufacturers’ IFU. Some sterilizer manufacturers’ IFU may include limitations on packaging.
AORN’s complete updated guideline is available for purchase at www.aorn.org. ✥
Susan Klacik, BS, CRCST, CIS, CHL, ACE, FCS, serves as a Clinical Educator for the International Association of Healthcare Central Service Materiel Management (IAHCSMM).
AORN. Guideline for sterilization packaging systems. Denver, CO: AORN, 2019. www.aorn.org.