Low-temperature sterilization: A disinfection alternative

By: OR Manager
Share

Some complex medical devices cannot withstand high-temperature sterilization. An alternative is low-temperature sterilization, which works for heat- and moisture-sensitive devices. There are different types of low-temperature sterilizers, and this article will review them.

 

Spaulding classification

A deciding factor in whether an instrument can undergo high-level disinfection is the Spaulding Classification, which identifies medical devices as critical, semicritical, or noncritical based on the risk of infection from contamination.

• Critical devices. These are introduced directly into the human body via the bloodstream or contact with other normally sterile areas of the body. These devices must be sterile.

• Semicritical devices. These contact intact mucous membranes or nonintact skin during use, but do not usually penetrate the blood barrier or other normally sterile areas. If a semicritical device cannot be sterilized, it must undergo high-level disinfection in which a sterilant is used but for a shorter exposure time than that required to achieve sterilization.

• Noncritical devices or instruments. These pose the lowest risk of transmission of infection, usually contacting only intact skin. These devices must be thoroughly cleaned and might require intermediate or low-level disinfection.

Regulatory bodies and organizations that set standards and guidelines have adopted the Spaulding classifications.

With terminal low-temperature sterilization, the item is packaged to protect it from contamination, and packaged instrumentation does not require processing just before use.

 

Choosing low-temperature sterilizers

Several different types of low-temperature sterilizers are available. A team of staff members from sterile processing, infection prevention and control, risk management, healthcare technology, education, and the OR should collaborate on deciding which type to purchase.

This team should review the medical devices being considered for low-temperature sterilization and check the manufacturer’s instructions for use (IFU) and validations for these products.

Medical devices are validated for repeated sterilization and reuse. The medical device manufacturer must demonstrate that the sterilization method in their IFU is compatible with the instrument’s materials, product configuration, and product functionality.

After being subjected to the sterilization method multiple times, the material is checked for physical strength, sterilization, physical dimensions, resilience, residues, and permeability (for ethylene oxide sterilization).

 

Low-temperature terminal sterilization

Ethylene oxide (EO), hydrogen peroxide gas plasma, vaporized hydrogen peroxide, and dual sterilants are the current terminal low-temperature sterilization methods. With all methods, instruments must be thoroughly cleaned before being sterilized.

Ethylene oxide

An EO sterilization cycle takes place in a sealed, temperature-controlled chamber. The process consists of air removal, conditioning, and injecting of EO; exposure to EO; removal of EO; and flushing, aerating, and air admission, allowing the opening of the chamber door.

EO sterilizers use a 100% EO cartridge, and the process takes approximately 16 hours.

Once the sterilization cycle starts, the chamber is heated and a vacuum is created to remove air from the chamber. Next, the conditioning phase treats the load before EO is injected to reach a predetermined temperature and relative humidity. The EO is injected into the chamber in a vaporized form, and then the exposure phase begins with a preset time.

At the end of the exposure time, the EO is removed and flushed from the chamber. The chamber is repeatedly filled with filtered air and cleared by a vacuum. During the aeration cycle, EO is removed from all items and packaging in the chamber. Finally, filtered air enters the chamber, bringing the chamber pressure equal to ambient pressure.

EO has exceptional penetration properties and vaporizes easily, so it can reach recessed areas and matted surfaces. More information is available in the ANSI/AAMI Standard ST41: Ethylene oxide sterilization in health care facilities.

Hydrogen peroxide gas plasma

Hydrogen peroxide gas plasma uses a combination of hydrogen peroxide vapor and low-temperature gas plasma for rapid sterilization. A proprietary cassette of hydrogen peroxide gas plasma is used to deliver the sterilant. This system is compatible with most materials and leaves no toxic residues. The load temperatures do not exceed 55°C, and sterilization occurs in a low-moisture environment. Standard cycle times are 28 and 38 minutes.

This system consists of two consecutive and equal sterilization phases. The sterilizer is loaded, the chamber door is closed, and pressure is further reduced.

The hydrogen peroxide solution is vaporized and transferred to the sterilization chamber. The hydrogen peroxide diffuses throughout the chamber surrounding the items to be sterilized and inactivates microorganisms. The pressure in the system is first increased and then, following a subsequent pressure reduction, applying energy to create an electric field generates low-temperature plasma, which in turn initiates the generation of the plasma.

In the plasma step, the hydrogen peroxide vapor is broken apart into reactive species known as free radicals. The plasma power is then terminated, causing the free radicals to lose their high energy and recombine to form oxygen, water vapor, and other nontoxic byproducts.

Repeating the above steps a second time completes the cycle. The sterilization process, therefore, consists of two consecutive, equal, and distinct half-cycles. When the second half-cycle is complete, the plasma power is turned off. The chamber is brought to atmospheric pressure by introducing HEPA-filtered air.

Vaporized hydrogen peroxide (VHP) uses hydrogen peroxide as the sterilant in a vaporized form. The load is processed through a short conditioning phase during which the chamber is evacuated to less than 1 Torr.

The next cycle is the conditioning phase, during which the load is checked for moisture content. If that is not acceptable, the load is aborted.

The next cycle reduces the pressure further, in preparation for injection of the VHP. The sterilizer automatically vaporizes the correct amount of liquid hydrogen peroxide, and then the sterilizer goes into the sterilization exposure phase.

After the sterilization phase, the load is aerated. The chamber VHP is exhausted through a catalytic converter that decomposes the VHP to water and oxygen.

No special venting is required for operation of the V-PRO 1 Sterilizer. When the sterilization cycle is complete, the load is removed and can be used immediately or stored for later use.

Total sterilization times range from 28 to 60 minutes depending on the cycle selected.

Dual-sterilant method

New this year is a system that uses dual sterilants: vaporized hydrogen peroxide and ozone for terminal sterilization of heat- and moisture-sensitive medical devices. It sterilizes using a proprietary sterilant at 105.8°F (41°C), and cycle time is between 46 and 70 minutes based on load composition and condition. There is only a single preprogrammed cycle for all loads, and the loads can be mixed.

A micro-pulsed hydrogen peroxide injection system provides sterilant until a predetermined pressure set point is reached, and then a small dose of ozone is injected into the chamber, scrubbing residual hydrogen peroxide from load contents. A chemical reaction forms hydroxyl radicals, adding to the lethality of the sterilization process. Near the end of the cycle, a catalytic converter breaks down the ozone and hydrogen peroxide into water vapor and oxygen. The sterilizer operator is notified when the cycle is complete.

 

Packaging

General packaging practices are followed for low-temperature sterilization, but other things must be taken into consideration. For each sterilization method, the packaging selected is based on the IFUs from the sterilizer manufacturer. The package must also be validated for that specific sterilization method, and the IFUs of the packaging manufacturer must be followed.

In addition, packaging for EO sterilization must allow for penetration of EO and aeration to remove all residual EO from the contents and packaging. For hydrogen peroxide sterilization, packaging must be cellulose free and allow for the deep vacuum draw without absorbing the sterilant, interrupting the cycle, or damaging the contents.

 

Quality monitors

Quality monitors are used to verify that sterilization parameters have been met. Low-temperature sterilization uses the same types of monitors as those used for steam sterilization. Each type of quality monitor must be validated for the specific type of sterilization. Internal and external chemical indicators are used.

The sterilizer operator must verify that parameters have been reviewed and met. The IFUs from the biological monitor must be adhered to, and the biological monitor must be validated for the specific type of sterilization.

It is important to review the sterilizer IFUs as they may describe routine or daily maintenance that needs to be performed.

 

IFU adherence

When deciding on a low-temperature sterilization method, it is important to check the IFU from the medical device manufacturer. Then review other critical aspects of sterilization to ensure they are validated for that method.

Staff training and competency are paramount to ensure proper sterilization. Policies and procedures must be developed following best practices and the IFUs from the sterilizer manufacturer. ✥

 

Klacik Sue 2014, low temperature sterilization, disinfection alternativeSusan Klacik, BS, CRCST, CIS, ACE, FCS, is president, Klacik Consulting (klacik.consulting.com), in Canfield, Ohio, and CSS manager at St Elizabeth Health Center and the IAHCSMM (International Association of Healthcare Central Service Materiel Management) representative to AAMI (Association for the Advancement of Medical Instrumentation).

References

Association for the Advancement of Medical Instrumentation. Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers. AAMI TIR12: 2010. www.aami.org.

Association for the Advancement of Medical Instrumentation. Chemical sterilization and high-level disinfection in health care facilities. ANSI/AAMI ST58: 2013. www.aami.org.

Association for the Advancement of Medical Instrumentation. Ethylene oxide sterilization in health care facilities: Safety and effectiveness. ANSI/AAMI ST41: 2008. www.aami.org.

Association for the Advancement of Medical Instrumentation. Flexible and semi-rigid endoscope processing in health care facilities. ANSI/AAMI ST91: 2015. www.aami.org.

International Association of Healthcare Central Service Materiel Management. Central Service Technical Manual, 8th ed. Chicago: IAHCSMM, 2016.

Join our community

Learn More
Video Spotlight
Live chat by BoldChat