The new year brings a newly updated Guideline for Manual Chemical High-Level Disinfection (formerly the Guideline for High-Level Disinfection). This AORN guideline, which is based on a comprehensive, systematic review of research- and non-research-based evidence, recommends best practices for safe and effective high-level disinfection (HLD) performed manually when automated methods are not possible.
Reusable semicritical items that have been validated by the manufacturer for sterilization should be sterilized if possible because HLD poses a greater risk of disease transmission than items processed by sterilization. Most HLD is performed in the preferred method of an automated processor. Automated methods—such as automatic endoscope reprocessors (AERs) for HLD or sterilization of reusable semicritical items—improve cleaning effectiveness, increase efficiency, and minimize personnel exposure to hazardous chemicals.
Equipment that is sterilized using automated methods can be better monitored for quality and consistency than when manual methods are used. However, some medical devices cannot undergo HLD in an automated processor, and this new guideline explains ways to safely and effectively perform manual HLD. The purpose of this guideline is to provide direction for:
• performing safe and effective manual chemical HLD of reusable semicritical items
• preventing patient and healthcare worker injury associated with the handling and use of liquid chemical high-level disinfectants.
When selecting a disinfection or sterilization method, it is important to review all available options. Sterilization provides the greatest margin of safety. An interdisciplinary team should perform a risk assessment to review semicritical devices that secondarily enter sterile tissue or the vascular system for sterilization or HLD.
There are 12 guideline recommendations that discuss how to safely and effectively preform manual HLD—notably by thoroughly cleaning the device and following manufacturer’s instructions for use (IFU).
The area where HLD is performed should be controlled and maintained to prevent contamination of the environment, cross contamination, and employee exposure while improving efficiency and enhancing process control and monitoring. Warning signs may be posted at the entrance to alert staff about the potential for exposure to hazardous chemicals.
HLD should occur in a room that is separate from patient care and not in high traffic areas. There should be two rooms, one for cleaning and one for HLD. Having two separate rooms reduces the chance of device contamination after disinfection because droplets and aerosols created during cleaning can cause cross contamination of nearby clean devices or surfaces.
If the two-room configuration is not possible, and the cleaning and disinfection are done in one room, the areas for HLD should be separated from the decontamination area by at least 3 feet, and there should be a separating wall or barrier that extends a minimum of 4 feet above the rim of the sink. To prevent cross-contamination, there should be a one-way flow of items undergoing processing. Having adequate space for staff to perform their tasks will decrease the chance for spills.
Not all high-level disinfectants are effective against certain types of viral pathogens. For example, research shows both glutaraldehyde and orthophalaldehyde do not demonstrate any significant reduction of infectious human papillomavirus type 16. These types of high-level disinfectants are commonly used to disinfect vaginal and rectal endocavity ultrasound probes.
In addition, inadequately disinfected endocavity probes increase the risk for transmission of HIV, hepatitis B virus, hepatitis C virus, Neisseria gonorrhea, Chlamydia trachomatis, and Trichomonas vanginalis. Even if sheaths are used, these probes must be thoroughly cleaned and disinfected. A system that uses a nebulized hydrogen peroxide mist method is more effective than HLD.
Interdisciplinary teams should be established to select the type of high-level disinfectants that will be used and decide how to store them. Using specially designated storage locations will reduce the possibility of spills and personnel exposure. Activated high-level disinfectants should be kept in tightly closed containers that are labeled with the contents and re-use life dates.
Point-of-use cleaning and proper transport must occur as medical devices are sent to the decontamination room for cleaning. All medical devices must be thoroughly cleaned because high-level disinfectants are inactivated or are less effective in the presence of organic material.
In addition, some high-level disinfectants act as fixatives and may fix organic soil and blood to the surface of the item being disinfected.
After cleaning, rinse the item and flush any lumens or channels with water in accordance with the device manufacturer’s IFU. The device manufacturer may recommend a final rinse with water treated to remove organic and inorganic substances, which will prevent staining and help ensure effective HLD or sterilization.
Following the manufacturer’s IFU for HLD ensures the effectiveness of the disinfection process or the performance of the device. The medical device manufacturer has validation performed to ensure the device can be effectively disinfected, and provides validated instructions for HLD.
Testing must be done to verify that the HLD process is effective for use—for example, by using a test strip or some other testing device cleared by the Food and Drug Administration that is specific to the disinfectant and the active ingredient in the disinfectant. Use a calibrated thermometer to verify the temperature and a timer to verify the exposure time.
The soaking container should be large enough to completely immerse the item while reducing the surface contact area to the smallest amount possible. Using the smallest container possible helps reduce the amount of high-level disinfectant that is needed and decreases the potential for exposure and spills.
Multipart instrumentation should be disassembled, and the item should be completely immersed in the HLD solution. Moving parts should be actuated while immersed to ensure all surfaces are exposed to the high-level disinfectant.
Lumened items should be flushed and completely filled with the high-level disinfectant. In addition, a timer should be used to ensure the item was exposed to the full soaking time as required in the manufacturer’s IFU.
After disinfection, the device should be safely and thoroughly rinsed with sterile water. Rinsing is extremely important to remove the high-level disinfectant. Thorough rinsing and flushing with sterile water helps prevent patient injury associated with any disinfectant remaining on the device. Rinse sinks should be located as close as possible to the high-level disinfectant soaking container.
After rinsing, if the device will be stored for future use, it should be dried using a clean lint-free cloth, and lumens should be flushed with instrument air.
After disinfection is complete, the device should be protected from contamination and clearly identified as processed and ready for use. This will help to prevent confusion and the accidental use of a contaminated device. Guidance is provided for protecting disinfected items from contamination and for transferring them onto a sterile field.
An entire section of the guidelines is dedicated to providing a safe environment for personnel who handle or use high-level disinfectants. Included is a discussion of safe high-level disinfectant handling, and a thorough spill control plan is presented describing how to prepare for a spill along with the regulatory requirements.
Maintaining records is vital, and there is guidance on what to record when performing HLD. Proper documentation will identify trends and show compliance with regulatory requirements and accreditation agency standards.
As with all medical device processing activities, education and training are paramount to a having a quality program. Recommendation X in the new guideline states: “The healthcare organization should provide initial and ongoing education and competency verification activities. It is the responsibility of the healthcare organization to provide initial and ongoing education and to verify the competency of its personnel; however, the primary responsibility for maintaining ongoing competency remains with the individual.”
The International Association of Healthcare Central Service Materiel Management has launched a scope certification program that fits well into this recommendation, along with a new endoscope reprocessing manual and workbook. For information, visit www.iahcsmm.org.
A quality assurance and performance improvement program can make it easier to identify problem areas, evaluate and improve the quality of patient care, and plan for corrective action. This section provides guidance on what to include in a quality management program.
A periodic review and evaluation of manual chemical HLD is needed to verify compliance or identify areas for improvement. Monitoring compliance with established safe practices can increase workplace safety and help reduce personnel exposure to high-level disinfectants. All staff who work with high-level disinfectants should participate in the quality assurance and performance improvement program.
Providing patients with high-quality medical devices is the goal of this new AORN guideline. Individual references have been appraised and scored, and the recommendations are rated according to the strength and quality of the evidence supporting each one. The recommendations, which are intended to be achievable, represent what is believed to be an optimal level of practice.