The evidence base consists of a single controlled study of spinal fusion that compared the accuracy of pedicle screw placement in a patient group that had two-dimensional (2-D) and three-dimensional (3-D) intraoperative imaging using the O-arm System (187 screws in 37 patients) and a patient group that had 2-D intraoperative fluoroscopy with postoperative computed tomography (152 screws in 30 patients).
The included studies have several limitations. Protocols for evaluating pedicle screw placement vary. Studies assessed use of O-arm in conjunction with image guidance. Determining how integration with a surgical navigation system affects imaging and procedure time was not possible. No comparative study reported on important outcomes of interest (i.e., total radiation exposure times, revision surgery rates). Researchers based comparative accuracy outcomes on radiographic metrics. Since the significance of cortical breaches in the absence of clinical symptoms remains uncertain, applicability of the findings to clinical practice is unclear.
Assessment of consistency for most outcomes was not possible because only one study met our inclusion criteria for efficacy. All case series we assessed reported no adverse events directly associated with full-rotation three-dimensional intraoperative imaging during spinal procedures. Pedicle misplacement rates in studies varied ranging from 0% to 15.6%.
| SUPPLEMENT TO OR MANAGER VOL 27, NO 11, NOVEMBER 2011 www.ormanager.com | ECRI Institute 5200 Butler Pike Plymouth Meeting, PA 19462-1298, USA Tel +1 (610) 825-6000, Fax +1 (610) 834-1275, Web www.ecri.org |