A regular column on sterilization and infection control issues.

When you sterilize an instrument set, you know what the standards are. Sterilizers must provide a sterility assurance level (SAL) of 10 to the minus 6 (10-6). Biological indicators are used to test the efficacy of the sterilizer. Chemical indicators or chemical integrators are placed in instrument sets to allow an OR nurse to determine whether an instrument set has been exposed to the sterilization process.
But there is no standard for measuring cleanliness. There is no cleaning monitor you can place in an instrument set. You can only determine whether something is cleaned by visual inspection, which may vary depending on your eyesight, available light, and whether a lighted magnifying glass was used. Products are available to test the efficacy of automated washing systems. But many instruments cannot be cleaned in automated systems, and some facilities, particularly small ambulatory surgery centers, do not have automated washing systems.
Yet effective cleaning is a prerequisite to effective sterilization or disinfection. Ineffective cleaning can compromise sterilization or disinfection and has been associated with adverse patient outcomes. Because there is no objective measurement to define when a device is clean, the best guarantee is consistent and strict adherence to:
Cleaning is defined by AAMI as the removal of contamination from an item to the extent necessary for further processing or for the intended use. AAMI notes that:
“…cleaning consists of the removal, usually with detergent and water, of adherent soil (eg, blood, protein, and other debris) from the surfaces, crevices, serrations, joints and lumens of instruments, devices and equipment by a manual or mechanical process that prepares the items for safe handling and or further decontamination.”
Cleaning removes microbial contamination and organic and inorganic material, which if not removed, can interfere with sterilization or disinfection.
Here is advice to help ensure your organization is taking the steps needed for effective cleaning.
Cleaning should begin as soon as possible after a device is contaminated.
Cleaning begins in the operating room, and steps to promote removal of contamination may begin during a procedure.
Some key points:
Instruments should be cleaned according to an established protocol by personnel who have demonstrated competency in instrument cleaning.
In an ideal world, soiled instruments would be immediately transported to the decontamination area, where they would be received by a certified instrument processing technician wearing personal protective equipment (PPE), who would immediately begin the decontamination process. The technician would determine which instruments must be manually cleaned and appropriately prepare and place all others in an automated system.
In reality, instruments may sit for a long time before they can be cleaned because of work loads, instrument turn-around priorities, and possible unavailability of automated washing systems. For this reason, instruments should be kept moist, either by adding a moistened towel to the transport container or applying a foam, spray, or gel intended for this purpose.
If the delay is prolonged, such as over a weekend, use a precleaning disinfectant to prevent biofilm formation. A biofilm is a mass of bacteria tightly adhered to a surface and not easily removed. Although biofilms may form on many surfaces, they are more common in lumened devices. If a biofilm breaks loose within a patient, the infection consequences can be severe and can lead to death.
Automated cleaning is preferable to manual cleaning. Automated cleaning systems are consistently effective and reduce the risk of exposing personnel to microbial contamination. Automated systems include washer-disinfectors, washer-decontaminators, washer-sterilizers, sonic irrigators for lumened devices, and ultrasonic cleaners. Phases in automated systems may include rinsing, enzymatic soak, detergent wash, ultrasonic cleaning, chemical germicide rinse, lubrication, and drying.
When automated systems are used, the manufacturer’s instructions are paramount. The instructions help in determining the device’s compatibility with the intended cleaning process, the correct type and concentration of detergent, necessary water quality, temperature, time, and care and maintenance of the equipment.
Whenever possible, monitor the process with products designed for automated processes. There are products for monitoring the cleaning efficiency of washer-disinfectors and ultrasonic cleaners and for monitoring cleaning efficacy within lumens.
Though automated cleaning is preferred, manual cleaning is needed in some situations, such as when the instrument inventory is inadequate, and there is need for quick turnaround. Although manual cleaning poses the greatest risk for inadequate cleaning, if performed correctly, it can be as effective as automated cleaning. The keys are:
Here are some reminders on effective manual cleaning:
Cleaning instruments with detergent is not like washing dishes at home. The instructions for use and concentration must be followed precisely. Detergent that is too concentrated or diluted can compromise the sterilization or disinfection process.
Here are a few tips:

Contaminated gastrointestinal flexible endoscopes have been linked to infections. These devices should be cleaned according to the guidelines of the Society of Gastroenterology Nurses and Associ-ates (www.sgna.org).
Instruments that have been in contact with prions should be cleaned according to a specific prion deactivation protocol, as recommended by AORN. In developing a protocol, consult recommendations of the World Health Organization and the Centers for Disease Control and Prevention.
Fortunately the days are long gone when it was thought that anyone who knew how to wash dishes was competent to wash surgical instruments. All staff responsible for instrument cleaning should have demonstrated competence. One method for developing competencies is to group instruments into processing needs, such as lumened devices, devices with difficult-to-clean areas, lensed instruments, and powered instruments, and determine the necessary competencies for each.
Cleaning is an essential step in instrument reprocessing and must be done in accordance with accepted protocols. It is no longer acceptable to discover “sterile dirt” after an instrument set is opened.
–Cynthia Spry, RN, MA, MSN, CNOR
Independent Clinical Consultant
Selecting a detergent is not a simple task and should always be done in conjunction with the detergent manufacturer, considering:
Because hard water may cause staining of instruments, a water-softening system may be necessary. The mineral content of the water must also be factored in.
For automated washers, detergents with a high pH are most effective. They can, however, be harsh on devices made from materials other than stainless steel. A milder detergent may be used on softer metals such as aluminum and on glass.
Manual cleaning formulations generally have a neutral pH of between 7 and 9. A detergent with a neutral pH will minimize damage to soft metals but will not dissolve water-insoluble proteins, making brushing and mechanical friction especially important.
Nonenzymatic detergents take longer to dissolve proteins. Because enzymes are effective for breaking down organic substances, most facilities use a detergent containing one or more enzymes. For facilities that specialize in orthopedics, a detergent with a lipase enzyme, which is effective on fats and organic materials, may be most appropriate. For facilities that specialize in vascular procedures, a detergent with a protease enzyme may be most appropriate. Protease enzymes are effective on blood, mucus, and proteins.
Association for the Advancement of Medical Instrumentation. Comprehensive Guide to Steam Sterilization and Sterility Assurance in Healthcare Facilities. ANSI/AAMI ST79:2006. Arlington, Va: AAMI.
Association of periOperative Registered Nurses. Recommended practices for cleaning and caring for surgical instruments and powered equipment. In Standards, Recommended Practices, and Guidelines. Denver: AORN, 2007.
Chu N S, Favero M S. Cleaning: An important prerequisite for instrument sterilization and disinfection. Infect Control Today. August 2001.