An occasional column on Joint Commission issues.
The reprocessing and storage of laryngoscope blades has been a hot button for Joint Commission surveyors, and OR managers have been asking how they can comply with the expectations.
In answer, the Joint Commission on October 24, 2011, issued a response to a frequently asked question (FAQ) on the issue.
The FAQ refers to the 2003 CDC/HICPAC Guidelines for Preventing Health-care Associated Pneumonia from the Centers for Disease Control and Prevention’s Healthcare Infection Control Practices Advisory Committee (www.cdc.gov/mmwr/preview/mmwrhtml/rr5303a1.htm).
Per Joint Commission standard IC.01.05.01 EP 1, organizations must follow “evidence-based national guidelines” such as these when developing infection prevention and control activities.
As the FAQ notes, the very last page of the CDC/HICPAC guidelines lists laryngoscope blades as semicritical devices. If you search for the word “semicritical” (no hyphen), you’ll find the following requirement (please note, it’s a Category 1A, the strongest of all recommendations): “Whenever possible, use steam sterilization (by autoclaving) or high-level disinfection by wet heat pasteurization at >158ºF (>70ºC) for 30 minutes for reprocessing semicritical equipment or devices.… After disinfection, proceed with appropriate rinsing, drying, and packaging, taking care not to contaminate the disinfected items in the process (308;310). Category IA.” The italicized section highlights a key issue that Joint Commission and Centers for Medicare and Medicaid Services surveyors are looking for specifically: wrapping of laryngoscope blades in individual packaging to prevent contamination while being stored for the next use.

During a survey, the trigger for recommendations often involves the surveyor’s inspection of equipment drawers in anesthesia carts and discovery of a collection of unwrapped blades in these locations.
Application of the CDC’s recommendation would be to clean the blade, subject it to high-level disinfection (HLD) or sterilization, and then ensure the blade is packaged and stored in a manner that prevents contamination.
An example is storing the blades in a clean sandwich-type bag or even a Ziploc-type bag. (Note: Neither the CDC nor Joint Commission has specified that such bags need to be sealed; the bag simply needs to cover or protect the blade from coming into contact with blood, body fluid, or organic material.)
A suggested laryngoscope disinfection and storage policy might include the following.
Infection Control Policy: High-Level Disinfection of Laryngoscope Blades
Laryngoscopes blades, non-CSR* reprocessing:
Laryngoscopes reprocessed in CSR*:
Testing laryngoscope light source/handle and blade
*CSR = Central sterile reprocessing.
This suggested policy is provided by Russell N. Olmsted, MPH, CIC, president, Applied Epidemiology Solutions, Inc, Ann Arbor, Michigan. Olmsted is a subcontractor to Patton Healthcare for issues related to infection prevention and control.
The rationale is that the blade is a semicritical device and needs to undergo appropriate reprocessing and be protected from contamination prior to the next use. These blades are also used in a variety of other locations, such as the emergency department, labor & delivery, NICU, difficult airway carts/boxes, and crash carts, so we encourage managers to look at these other locations in addition to the surgery suite.
Some areas may choose to up-grade from the minimum of high-level disinfection to steam sterilization–but the minimum standard is HLD. Some may decide to use sterilization because the sterilized blades are then wrapped and protected. This may be optimal but is not required.
If the blade is HLD and contained in a plastic bag, and testing of the light source/blade is desired, it may be possible to manipulate the blade onto the light source/handle and test it without actually removing the blade from the bag and without actually touching the blade itself.
Even if the blade has been sterilized in a peel pouch, and light source/blade testing is desired, the provider is likely to be able to partially remove the blade from the peel pouch, test it, and slip it back into the peel pouch for on-going protection. In either case, proper hand hygiene practices should be observed prior to handling the blade.
One other tip: review the instructions for use (IFU) from the manufacturer of the blade light source/handle. This component is not semicritical but is a noncritical device. As such, it does not need to undergo HLD or sterilization based on CDC recommendations. However, some manufacturers’ IFUs do specify disinfection between patient uses, so review this information for the light source/ handles in use at your facility.
A suggested policy for laryngoscope disinfection and storage is in the sidebar.
—John Rosing, MHA, FACHE
Vice President and Principal
Patton Healthcare Consulting
Thiensville, Wisconsin
John Rosing can be reached at 262/242-3631 or johnrosing@pat-tonhc.com.
The Joint Commission FAQ is at www.jointcommission.org/standards_information/jcfaq.aspx. The FAQ applies to hospitals, critical access hospitals, and ambulatory care organizations.