The use of electrosurgical instrumentation is common in minimally invasive surgery (MIS). However, while these instruments are common, processing them is complex and time-consuming. Insulated electrosurgical instrumentation can cause patient harm, including an electrical burn, if there is a breach in the insulation.
These instruments are designed to have an electrical current flow from the current source to the patient. They are generally used for cutting, dissecting, and cauterizing. The insulation keeps the electrical current from contact with the patient until it reaches the exact surgical site.
A break in the insulation can also result in debris getting under it, which can go undetected; therefore, it must remain completely intact. Non-intact insulation may cause arcing and ricocheting inside the body cavity. Often, the damage to surrounding organs cannot be seen by the OR team and may be left undiscovered until a patient returns with an infection or other complications.
There are three generations of laparoscopic instruments:
• The first is a one-piece instrument that cannot be disassembled and lacks a flushing port. This generation is the most difficult to clean because there is no access to the inner lumen for cleaning.
• The second has a flushing port to flush the lumened part of the instrument; however, it cannot be disassembled for cleaning.
• The third can be disassembled into three or four pieces, allowing for better cleaning and sterilization. This generation may or may not have a flushing port.
All three generations may be seen at one facility and even in the same set.
As with all instruments, the cleaning of laparoscopic devices should begin at the point-of-use. Laparoscopic instrument shafts should be wiped to remove gross soil. It is important that cannulated instruments are flushed at the point-of-use to prevent fluids from coagulating in the shaft because the dry soil becomes extremely difficult to clean.
Processing laparoscopic instruments is a complex process, and the manufacturer’s instruction for use (IFU) should be carefully followed during the cleaning process.
The outside of the instrument should be cleaned using a clean soft cloth. The distal tips, joint areas, and insulated shafts must also be carefully cleaned. The insulation is difficult to inspect for debris because its dark color can hide organic soil that may be present on the instrument shaft. To protect the integrity of the insulated shaft, it should be cleaned with a clean, soft lint-free cloth. Brushes used for cleaning the working tips should be nylon, not metal, because metal brushes may scratch the insulation.
Laparoscopic instruments that only have a flush port should be thoroughly flushed with a cleaning solution, followed by a thorough rinse. If the instrument can be disassembled for cleaning, it should be disassembled, flushed, then manually cleaned using the correct-sized cleaning brush. The use of an automatic lumen flusher assists in the flushing process.
When brushing the lumened part of the instrument, the cleaning brushes must go completely through the instrument shaft, cleaning all surfaces, and exit at the other end. Before bringing the brush back through the lumen, it should be cleaned to remove soil so not to contaminate the lumen.
It is important to use the correct-sized brush so that the entire surface is brushed. If the brush is too large, it will not fit into the lumen and may even scratch the lumen walls. If the brush is too small, it will not have complete contact with the lumen walls and will not clean them thoroughly. Upon completion of the brushing, the lumens should undergo a thorough rinsing.
After the instruments are manually cleaned, they can undergo ultrasonic cleaning. The use of an irrigating ultrasonic cleaner provides additional lumen flushing. When using an irrigating ultrasonic cleaner, it is important to ensure the instruments are correctly attached to the flush tubing. The sonic cycle should be run for the recommended amount of time.
After the sonic irrigation process, instruments may be placed into a washer-disinfector for further cleaning. Insulated instruments should be secured so they do not move freely during this process because the high impingement action may damage the insulation. All instruments should be in the open position during the mechanical cleaning process.
Some washer-disinfectors have manifolds with lumen flushing capability. A manifold system provides a mechanism for lumened instruments to connect to a type of irrigation system that will flush the lumened devices during the cycles. It is essential that instruments are correctly attached to the flush tubing. The final rinse should be performed using critical water, as described in the AAMI TIR 34—Water for the reprocessing of medical devices. After cleaning, the instruments should be dried using a clean non-linting cloth and instrument air.
The next step is to inspect the instrumentation. While all instruments are visually inspected for cleanliness, these instruments should undergo enhanced inspection with magnification, borescopes, or other inspection methods to verify cleanliness.
Borescopes may be used to visually check internal channels of instrumentation for cleanliness and/or damage following the borescope’s IFU. If any debris or residue is detected, it should be returned to the decontamination room to be thoroughly cleaned. No type of cleaning should be performed in the preparation and packaging area.
Most importantly, every time electrosurgical instruments are processed, they should also undergo insulation testing. Damage to insulated instruments can occur during normal use, processing, contact with sharp instruments, and high voltage use; repeated use/processing can also cause damage to the insulation.
To inspect the insulation, locate the metal collar at the distal tip. The insulation should fit tightly against the collar; this union should be snug, with no spaces visible. Next, grip the insulation and try to slide it back. If the insulation slides (moves), the instrument should be sent for repair.
There are several types of insulation testing equipment commercially available. They typically include various accessories to test specific instrumentation and cables/cords. The manufacturer provides specific information on how to perform electrosurgical insulation testing. Each is designed to detect small current leaks that can place patients at risk of significant harm.
Any damage to the insulation can compromise the safety of these instruments. Damaged instruments must be removed from service, and the damage should be addressed according to organizational policy and procedure.
According to the AORN Guideline for use of electrosurgical devices, an interdisciplinary team that includes members of the perioperative and sterile processing departments should determine a standardized communication strategy and action plan, to perform when an insulation failure is found during processing. The communication should include sterile processing, OR personnel, and physicians.
Processing personnel should receive education regarding the processing of laparoscopic instrumentation, and they should also receive education in the use of all testing equipment before use. Competency should be verified and documented before first using the insulation tester. Both the Association for the Advancement of Medical Instrumentation (AAMI) and AORN have recognized the risks associated with insulated laparoscopic instrumentation and have included guidance in their standards and guidelines.
In 2020, an amendment was added to the ANSI/AAMI ST79:2017 Comprehensive guide to steam sterilization and sterility assurance in healthcare facilities; Amendment A.2 Inspection of insulated instruments. The AORN Guideline for use of electrosurgical devices was also revised in 2020 and includes a guideline to visually inspect electrosurgical accessories for damage (eg, insulation breakage) before and after use and, if reusable, during reprocessing.
—Susan Klacik, BS, FCS, ACE, CHL, CIS, CRCST, AAMIF, is clinical educator at Healthcare Sterile Processing Association (HSPA) and its representative the Association of the Advancement of Medical Instrumentation (AAMI).
Association for the Advancement of Medical Instrumentation. ANSI/AAMI ST79:2017 & 2020 Amendments A1, A2, A3, A4 (Consolidated Text) Comprehensive guide to steam sterilization and sterility assurance in healthcare facilities.
Guideline for electrosurgical safety. AORN Recommended Practices and Guidelines. 2022.
International Association of Healthcare Central Service Materiel Management. Central Service Technical Manual 39. 7th ed. Chicago: IAHCSMM, 2007.
International Association of Healthcare Central Service Materiel Management. 8th ed. Chicago: IAHCSMM, 2016.
International Association of Healthcare Central Service Materiel Management. Instrument Resource Manual. 1st ed. Chicago: IAHCSMM, 2018.