Techniques for using sterilization containers, case-by-case basis

By: Susan Klacik, BS, FCS, ACE, CHL, CIS, CRCST, AAMIf
Share

Sterilization containers are used in many surgery departments. They offer many benefits, including providing an exceptional barrier to microorganisms, protecting instruments from damage, and eliminating torn wrappers. These benefits can only be realized if the sterilization containers are used properly, however.

Guidance for the use of sterilization containers is given in ANSI/AAMI ST79 Comprehensive guide to steam sterilization and sterility assurance in health care facilities and the AORN Guideline for Sterilization Packaging Systems. These guidance documents both recommend that sterilization containers are thoroughly cleaned between uses and inspected before use to ensure the latching mechanisms are secure for sterilization; gaskets are pliable and securely fastened (and free of breaks or cuts); and filter retention mechanisms and fasteners are secure. Some containers use a valve system that works freely.

Following the sterilization containers instructions for use (IFU) is strongly recommended in both guidance documents—and for good reason. Each sterilization container has its own unique characteristics and performance requirements that are found in their IFU. Diligently following the IFU is imperative because those instructions are based on a scientific validation process that is reviewed by the US Food and Drug Administration, then cleared before being purchased by healthcare organizations.

There are many differences between sterilization containers such as design, size, material, mechanic, and construction—all of which affect the performance of the sterilization container. Some IFU from the same manufacturer may differ between different-sized containers.

What follows is critical information to look for in sterilization container IFU:

• sterilization modality or method of sterilization

• sterilization cycle (not all sterilization containers can be used in both gravity and dynamic air removal cycles; even though the IFU states steam sterilization, it may specify which cycle to use, and even specific parameters may be cited)

• lumen specifications (the IFU may state lumen size restrictions or the number of lumens per container)

• types of instruments and placement of instruments

• types of materials that can or cannot be used (some sterilization container IFU limit textile use)

• weight limits

• the most challenging area of the sterilization container for placement of chemical indicators (and if performing verification testing biological indicators)

• stacking restrictions

• specific inspection instructions

• cleaning chemicals to use (and in some cases, cleaning chemicals to avoid)

• routine maintenance to be performed.

In addition to checking the IFU of the sterilization container manufacturer, the IFU of the sterilizer should be consulted. There may be weight limits per load; some low-temperature hydrogen peroxide sterilizers limit the number of containers per load. Low-temperature hydrogen peroxide sterilization processes are based on a fixed amount of sterilant for each cycle type and for every load placed in the chamber. This fixed amount of sterilant is delivered in different manners, such as an ampule contained in a cassette or cup, depending upon the type of sterilizer used. It is important to note that by having a specific fixed amount of sterilant, there is little room for error since no additional sterilant is added during the sterilization exposure phase. All loads use the same amount of sterilant during the sterilization exposure phase. Overloading the sterilizer may result in a cancelled cycle or sterilization failure.

During use of sterilization containers, it is not unusual for questions or concerns to arise. It is best to contact the sterilization container manufacturer for direction. To avoid confusion and misunderstanding, this type of communication should be in writing and kept on file. The same holds true if questions or concerns arise regarding the use of sterilization containers with a sterilizer. Such questions should be directed to the sterilizer manufacturer. ORM

 

 


January 1, 2022, officially marked the transition from the International Association of Healthcare Central Service Materiel Management (IAHCSMM) to the Healthcare Sterile Processing Association (HSPA). The Association’s support and services will remain, and our new name will help improve understanding of the discipline, and its relevance in the healthcare community and the general public. For more information about HSPA, visit www.myhspa.org.